Skip to the content

FDA Medical Device Recalls (openFDA)

Medline Industries, LP: DEXLOCK Achilles Repair Implant Kits, MAKT4520

Recall number
Z-0911-2026
Recalling firm
Medline Industries, LP
Product
DEXLOCK Achilles Repair Implant Kits, MAKT4520
Status
Open, Classified
Initiated
2025-11-11
Posted
2025-12-11
Terminated
not on file
Reason
There have been multiple complaints of drill bit fusing to the bushing/sleeve within the drill guide during use. Retrieving a replacement drill bit and drill guide may prolong the surgical procedure.
Root cause
Under Investigation by firm
Action
Firm began notifying consignees on 11/11/2025 via Urgent Medical Device Recall letters. Customers have been instructed to over-label kits with a sticker instructing them to not use the drill guide within the Achilles Kit and to use the reusable 3.6mm drill guide from the Soft Tissue Instrument Tray. Customers who do not have a Soft Tissue Instrument Tray have been instructed to notify Medline so that a tray can be provided. Customers were instructed to quarantine all affected product in stock and return a completed response form to Medline. Upon receipt of the response form, the firm will then provide over-labels to place on affected inventory. If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.
Quantity
163 US
Distribution
US Nationwide distribution in the states of AL, AR, AZ, CO, DE, FL, GA, ID, IL, IN, KS, KY, LA, MA, MT, NC, NJ, OH, OK, PA, SC, TX, VA, WI.
Lot, serial or model codes
Lot 19019 UDI-DI 10193489131703
510(k) numbers
not on file
PMA numbers
not on file
Product code
NWN
Firm FEI
1417592
Firm city
Northfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register