Orthofix, Inc: Orthofix ISKD Limb Lengthener. ISKD - Intramedullary Skeletal Kinetic Distractor. The ISKD is intended for limb lengthening of the tibia and femur.
- Recall number
- Z-0895-2013
- Recalling firm
- Orthofix, Inc
- Product
- Orthofix ISKD Limb Lengthener. ISKD - Intramedullary Skeletal Kinetic Distractor. The ISKD is intended for limb lengthening of the tibia and femur.
- Status
- Terminated
- Initiated
- 2012-08-06
- Posted
- 2013-02-28
- Terminated
- 2013-05-17
- Reason
- There is a possibility that the ISKD limb lengthener may stop distracting post-operatively during treatment, which may result in premature bone consolidation (limb not achieving the desired length) leading to revision surgery to remove and/or replace the device.
- Root cause
- Nonconforming Material/Component
- Action
- Orthofix sent an Urgent Product Recall Notification letter dated August 6, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to discontinue distribution of devices, identify the affected and remove from their inventory, and returned to the firm. Credit will be issued upon receipt of the returned product. Customers were asked to take the actions listed in the enclosed Product Return Instructions immediately. Customers with questions should contact Orthofix Customers Service by telephone at 800-266-3349 or 214-937-2047. For questions regarding this recall call 214-937-2509.
- Quantity
- 1224
- Distribution
- Worldwide Distribution - USA (nationwide) Internationally to Australia, Hong Kong, Saudi Arabia, Korea, Turkey, Greece, Finland, South Africa, Spain, Singapore, Switzerland, Brazil, Germany, France, Japan, Netherlands, New Zealand, Israel, Canada and Mexico
- Lot, serial or model codes
- All serial numbers of the following Catalog Numbers: F12-255-305, F12-345-395NS, T12-245-295, F12-255-305NS, F12-345-425, T12-255-305, F12-255-335, F12-345-425NS, T12-255-305NS, F12-255-335NS, T10-215-265, T12-255-335, F12-300-350, T10-255-305, T12-255-335NS, F12-300-350NS, T10-255-335, T12-300-350, F12-300-380, T10-300-350, T12-300-350NS, F12-300-380NS, T12-215-265, T12-300-380, F12-345-395, T12-215-265NS, and T12-300-380NS
- 510(k) numbers
- ["K010322","K031219"]
- PMA numbers
- not on file
- Product code
- HSB
- Firm FEI
- 2183449
- Firm city
- Lewisville
- Firm state
- TX
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28