Skip to the content

FDA Medical Device Recalls (openFDA)

Waldemar Link GmbH & Co. KG (Mfg Site): Endo Model Knee Fusion Nail SK, Knee Fusion Nail, Coupling Component, right, modular, L= 68 mm, Taper 10/12. Knee component of limb salvage replacement devices.

Recall number
Z-0890-2026
Recalling firm
Waldemar Link GmbH & Co. KG (Mfg Site)
Product
Endo Model Knee Fusion Nail SK, Knee Fusion Nail, Coupling Component, right, modular, L= 68 mm, Taper 10/12. Knee component of limb salvage replacement devices.
Status
Open, Classified
Initiated
2025-11-10
Posted
2025-12-09
Terminated
not on file
Reason
The device has been delivered with a preassembled locking screw that had been inserted from the wrong side into the male taper of the femoral component.
Root cause
Process control
Action
On November 10, 2025, URGENT: MEDICAL DEVICE RECALL letters were emailed to distributors. Actions to be taken: Please respond using the attached Distributor Reply form to indicate your understanding of this Device Recall and of the need to return the affected product. Please return the product as soon as possible to the following address: LinkBio Corporation 69 King St Dover, NJ 07801 Please contact LinkBio at 973-625-1333 or complaints@linkbio.com for any questions.
Quantity
18 units
Distribution
US distribution to states of: AL, GA, and TX.
Lot, serial or model codes
Item Number: 15-0028/08; UDI-DI: 04026575175222; Lot Number: 2535004;
510(k) numbers
["K151008"]
PMA numbers
not on file
Product code
KRO
Firm FEI
3003386935
Firm city
Norderstedt
Firm state
not on file
Firm country
Germany

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register