GE Healthcare, LLC: SIGNA Pioneer Nuclear Magnetic Resonance Imaging System - Product Usage: is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times.

Recall number
Z-0876-2020
Recalling firm
GE Healthcare, LLC
Product
SIGNA Pioneer Nuclear Magnetic Resonance Imaging System - Product Usage: is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times.
Status
Terminated
Initiated
2019-12-23
Posted
2020-01-28
Terminated
2020-12-11
Reason
It was identified that due to a potential installation workflow issue, the MR system date could be set incorrectly. The system's date and time setting are used to populate the DICOM header information on images. This could result is an inaccurate date recorded on the images.
Root cause
Software design
Action
This issue will be resolved by sending an Urgent Medical Device Correction letter to Medical Facility staff (Director of Clinical/Radiology, Risk Manager/Hospital Administrator, Director of Biomedical Engineering) with affected systems. This letter will inform the customer of the issue and instruct the user to ensure the displayed system date is correct. The long-term solution is to correct affected systems with updated software.
Quantity
62 (30 US, 32 OUS) total devices
Distribution
Worldwide distribution - US nationwide distribution and countries of Australia, China, France, Germany, India, Italy, Japan, Netherlands, Oman, Spain, Sweden, Switzerland, Turkey, and United Kingdom.
Lot, serial or model codes
Mfg. Lot or Serial # System ID To be provided GON4833100 01264 860536MR3T KZTBL1700064TJ 860871MR3T KZTBL1800069TJ RADNETMR1700 KZTBL1800022TJ 508872PMR 01302 RADNETMR1708 KZTBL1900130TJ 301698MR1 KZTBL1900071TJ 704333PIO KZTBL1900110TJ 631751MR KZTBL1900107TJ RADNETMR1707 KZTBL1900121TJ 803936LMCMR2 033-4-1532 281HMI750W To be provided 83199160995 To be provided M24313812 To be provided M2850013 To be provided TP0013MR01 To be provided PNUE04MR02 To be provided TP0008MR07 To be provided 083027823408018 To be provided 083027823888617 To be provided GON4826677 To be provided GON4762453 To be provided B5142305 To be provided ZC0002 To be provided ZC0034 To be provided ZC0001 To be provided ZC0306 To be provided GON4795546 To be provided 222768MR01 To be provided 34340MRS02 To be provided 10627MRS02
510(k) numbers
K143345, K153003, K160621
PMA numbers
not on file
Product code
LNH
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register