Skip to the content

FDA Medical Device Recalls (openFDA)

Steris Corporation Hopkins Facility: STERIS SYSTEM 1E Liquid Chemical Sterilant Processing System. Model P6500, 510 (k) Numbers: K090036 Cleared 4/05/2010; K101409 cleared 8/2/2010; K102462, cleared 9/2/2010; add to file 2/24/2011. The SYSTEM 1E Liquid Sterilant Processing System is intended for the liquid chemical sterilization of manually cleaned immersible and reuseable critical and semi-critical heat sensitive medical devices, including endoscopes and their accessories.

Recall number
Z-0869-2012
Recalling firm
Steris Corporation Hopkins Facility
Product
STERIS SYSTEM 1E Liquid Chemical Sterilant Processing System. Model P6500, 510 (k) Numbers: K090036 Cleared 4/05/2010; K101409 cleared 8/2/2010; K102462, cleared 9/2/2010; add to file 2/24/2011. The SYSTEM 1E Liquid Sterilant Processing System is intended for the liquid chemical sterilization of manually cleaned immersible and reuseable critical and semi-critical heat sensitive medical devices, including endoscopes and their accessories.
Status
Terminated
Initiated
2011-12-07
Posted
2012-01-25
Terminated
2014-05-13
Reason
Software deficiency. STERIS identified that water specifications were not being met at some of their customer sites. Critical parameters for initiating cycles are: water quality, water pressure and water temperature. Failure to meet these criteriae may lead to cycle aborts by the SYSTEM 1E Liquid Chemical Sterilant Processing System. The firm intends to perform a software field correction which
Root cause
Software design
Action
The firm, STERIS Corp. prepared an "URGENT VOLUNTARY FIELD CORRECTION NOTICE" letter dated December 8, 2011. The firm intends to send the letter to its customers upon STERIS' receipt of 510(k) clearance from FDA. The letter includes a description of the Product-SYSTEM Liquid Chemical Sterilant Processing System, problem and actions taken. The customers were instructed to in case of an incomplete or cancelled cycle, reprocess devices in the cycle following the directions in the Operator manual. The letter also states that STERIS Corporation will install updated software on the affected system, and will be providing the customers with a new Operator Manual reflecting the revised wording resulting from the installation of the new software. For further information or if they have any questions regarding the firms visit to their facility, contact STERIS Field Service Dispatch at 1-800-333-8828. If you have questions regarding this matter, contact Director Low Temperature Sterilization, at 440-392-7455.
Quantity
3,882 units were distributed
Distribution
Worldwide distribution: USA (nationwide) and countries of: Bahrain, Hong Kong, Japan and South Korea.
Lot, serial or model codes
MED. Serial Numbers: 400000 through 405499 are subject to correction.
510(k) numbers
K090036, K102462
PMA numbers
not on file
Product code
MED
Firm FEI
3000251274
Firm city
Mentor
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register