Skip to the content

FDA Medical Device Recalls (openFDA)

Zimmer Inc.: Zimmer brand fracture blade plate instrument Chisel, 95 degree, small; assembly part 4705-03-10, catalog number 00-4705-003-10.

Recall number
Z-0864-05
Recalling firm
Zimmer Inc.
Product
Zimmer brand fracture blade plate instrument Chisel, 95 degree, small; assembly part 4705-03-10, catalog number 00-4705-003-10.
Status
Terminated
Initiated
2005-02-09
Posted
2005-06-02
Terminated
2005-06-29
Reason
The engraved calibration lines are incorrectly specified, which will result in the user obtaining a reading that the chiseling removed material to a depth 10mm deeper than what was actually removed, thus guiding the user to select a blade plate that is longer than will fit.
Root cause
Other
Action
The firm's distribution network was notified via email on 2/9/05, instructed to check their inventory and that of their customer hospitals, and to return the product to Zimmer.
Quantity
not on file
Distribution
Nationwide and Canada.
Lot, serial or model codes
Chisel lot number 60048084.
510(k) numbers
not on file
PMA numbers
not on file
Product code
LXH
Firm FEI
1000220733
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register