Medtronic Neuromodulation: Medtronic Intellis Spinal Cord Stimulator, Model Number A710, Clinical Programmer application version 1.3.80 - Product Usage: intended for use by clinicians in the programming of the following Medtronic neurostimulators (external and implantable) for pain therapy.

Recall number
Z-0860-2021
Recalling firm
Medtronic Neuromodulation
Product
Medtronic Intellis Spinal Cord Stimulator, Model Number A710, Clinical Programmer application version 1.3.80 - Product Usage: intended for use by clinicians in the programming of the following Medtronic neurostimulators (external and implantable) for pain therapy.
Status
Open, Classified
Initiated
2020-11-12
Posted
not on file
Terminated
not on file
Reason
A710 Intellis Clinician Application has a software issue that can result in the inability to program the Intellis implantable neurostimulation device.
Root cause
Software design (manufacturing process)
Action
Medtronic notified the first consignee by letter on 11/12/2020 with a confirmation of their complaint and request to perform a correction to the device which put the device in a factory-reset state and restore functionality to allow programming and continued therapy. On 17 Dec 2020, Medtronic will begin issuing an Urgent Medical Device Correction notification to physicians regarding this validation error and providing instructions on the impact and appropriate mitigating actions if this issue presents.
Quantity
7960 downloads
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Canada, Latin America, Asia Pacific, Japan, China, and European regions.
Lot, serial or model codes
Software version 1.3.80
510(k) numbers
not on file
PMA numbers
not on file
Product code
LGW
Firm FEI
2182207
Firm city
Minneapolis
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register