FDA Medical Device Recalls (openFDA)
Medtronic Neuromodulation: Medtronic Intellis Spinal Cord Stimulator, Model Number A710, Clinical Programmer application version 1.3.80 - Product Usage: intended for use by clinicians in the programming of the following Medtronic neurostimulators (external and implantable) for pain therapy.
- Recall number
- Z-0860-2021
- Recalling firm
- Medtronic Neuromodulation
- Product
- Medtronic Intellis Spinal Cord Stimulator, Model Number A710, Clinical Programmer application version 1.3.80 - Product Usage: intended for use by clinicians in the programming of the following Medtronic neurostimulators (external and implantable) for pain therapy.
- Status
- Open, Classified
- Initiated
- 2020-11-12
- Posted
- not on file
- Terminated
- not on file
- Reason
- A710 Intellis Clinician Application has a software issue that can result in the inability to program the Intellis implantable neurostimulation device.
- Root cause
- Software design (manufacturing process)
- Action
- Medtronic notified the first consignee by letter on 11/12/2020 with a confirmation of their complaint and request to perform a correction to the device which put the device in a factory-reset state and restore functionality to allow programming and continued therapy. On 17 Dec 2020, Medtronic will begin issuing an Urgent Medical Device Correction notification to physicians regarding this validation error and providing instructions on the impact and appropriate mitigating actions if this issue presents.
- Quantity
- 7960 downloads
- Distribution
- Worldwide distribution - US Nationwide distribution and the countries of Canada, Latin America, Asia Pacific, Japan, China, and European regions.
- Lot, serial or model codes
- Software version 1.3.80
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- LGW
- Firm FEI
- 2182207
- Firm city
- Minneapolis
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28