FDA Medical Device Recalls (openFDA)
Arizant Inc: Bair Hugger Temperature Management Blanket, Model 555 Pediatric Full Access Blanket. Use only with a Bair Hugger temperature managemen unit. Not Sterile. Do not use if primary packaging has been previously opened or is damaged. Arizant Healthcare Inc., 10393 West 70th Street, Eden Prairie, MN 55344 USA.
- Recall number
- Z-0855-05
- Recalling firm
- Arizant Inc
- Product
- Bair Hugger Temperature Management Blanket, Model 555 Pediatric Full Access Blanket. Use only with a Bair Hugger temperature managemen unit. Not Sterile. Do not use if primary packaging has been previously opened or is damaged. Arizant Healthcare Inc., 10393 West 70th Street, Eden Prairie, MN 55344 USA.
- Status
- Terminated
- Initiated
- 2005-03-11
- Posted
- 2005-06-01
- Terminated
- 2007-01-10
- Reason
- Air channel seals on a limited number of Bair Hugger Model 555 Pediatric underbody blankets could separate when inflated with air from Bair Hugger warming unit. This separation would create a bulging area in the blanket's air channels, which could have a potential to shift the patient's position on the blanket.
- Root cause
- Other
- Action
- Recall notification was mailed to customers beginning 03/11/05. Customers were informed of the issue and asked to check their inventory for the affected lots. Affected lots found are to be quarantined and customers are to contact Arizant for instruction on returning affected product and receiving replacement product. Distributors are instructed to examine their inventory for affected lots, discontinue distribution and quarantine the lots. Distributors are asked to contact Arizant for returning affected product. If affected product was distributed, Distributors are to contact Arizant with names, addresses and telephone numbers of their accounts so Arizant can immediately inform those accounts of the recall.
- Quantity
- 88,320 blankets
- Distribution
- nationwide
- Lot, serial or model codes
- lot numbers ranging from 9631 up to and including 9657
- 510(k) numbers
- K001149, K913734
- PMA numbers
- not on file
- Product code
- DWJ
- Firm FEI
- 2183725
- Firm city
- Eden Prairie
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28