Arizant Inc: Bair Hugger Temperature Management Blanket, Model 555 Pediatric Full Access Blanket. Use only with a Bair Hugger temperature managemen unit. Not Sterile. Do not use if primary packaging has been previously opened or is damaged. Arizant Healthcare Inc., 10393 West 70th Street, Eden Prairie, MN 55344 USA.

Recall number
Z-0855-05
Recalling firm
Arizant Inc
Product
Bair Hugger Temperature Management Blanket, Model 555 Pediatric Full Access Blanket. Use only with a Bair Hugger temperature managemen unit. Not Sterile. Do not use if primary packaging has been previously opened or is damaged. Arizant Healthcare Inc., 10393 West 70th Street, Eden Prairie, MN 55344 USA.
Status
Terminated
Initiated
2005-03-11
Posted
2005-06-01
Terminated
2007-01-10
Reason
Air channel seals on a limited number of Bair Hugger Model 555 Pediatric underbody blankets could separate when inflated with air from Bair Hugger warming unit. This separation would create a bulging area in the blanket's air channels, which could have a potential to shift the patient's position on the blanket.
Root cause
Other
Action
Recall notification was mailed to customers beginning 03/11/05. Customers were informed of the issue and asked to check their inventory for the affected lots. Affected lots found are to be quarantined and customers are to contact Arizant for instruction on returning affected product and receiving replacement product. Distributors are instructed to examine their inventory for affected lots, discontinue distribution and quarantine the lots. Distributors are asked to contact Arizant for returning affected product. If affected product was distributed, Distributors are to contact Arizant with names, addresses and telephone numbers of their accounts so Arizant can immediately inform those accounts of the recall.
Quantity
88,320 blankets
Distribution
nationwide
Lot, serial or model codes
lot numbers ranging from 9631 up to and including 9657
510(k) numbers
K001149, K913734
PMA numbers
not on file
Product code
DWJ
Firm FEI
2183725
Firm city
Eden Prairie
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register