Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: Artis zee/ zeego systems, model numbers: 10094135, 10094137, 10094139, 10094141, 10280959: dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed include cardiac angiography, neuro angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures.

Recall number
Z-0854-2015
Recalling firm
Siemens Medical Solutions USA, Inc
Product
Artis zee/ zeego systems, model numbers: 10094135, 10094137, 10094139, 10094141, 10280959: dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed include cardiac angiography, neuro angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures.
Status
Terminated
Initiated
2014-09-17
Posted
2015-01-14
Terminated
2015-08-06
Reason
The Small Focus of a 3 focal spot x-ray tube may fail and further release of radiation will not be possible. The Large Focus and Micro Focus focal spots will still be operational without any restriction. Manual switchover is needed, resulting in a short interruption of the procedure.
Root cause
Process control
Action
Planned action (s) to repair defect or to bring product into compliance for Siemens Medical Solutions Inc. concerning the corrective action plan (CAP) for the Model(s) Artis Zee/Zeego Stationary Angiographic X-ray Systems. 1.Notifying the customer of the defect. 2.Implementing a corrective software patch to protect the system from damage in the event of hardware failure 3. Scheduling a service visit from a Siemens representative to perform a hardware replacement. 4. Replacing the non-conforming X-ray tube with one manufactured after the nonconformity issue was resolved. For further questions please call (610) 219-6300.
Quantity
44
Distribution
USA Distribution to the states of MO, MI, NY, OK, CA, LA, CT, IN, VA, MD, WA, HI, ME, TX, NY, OH, PA and CO.
Lot, serial or model codes
model numbers: 10094135, 10094137, 10094139, 10094141, 10280959, with serial numbers: 160348 146643 147129 146630 147208 146972 153707 147378 146742 157650 146986 147098 157696 153946 146940 147144 147029 160384 153884 147009 153628 153706 146729 147155 153790 146896 153757 136503 147372 136345 153762 146883 160421 153195 146427 153834 135927 154157 135862 153998 153257 153220 153682 160341
510(k) numbers
["K073290"]
PMA numbers
not on file
Product code
IZI
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register