Skip to the content

FDA Medical Device Recalls (openFDA)

General Electric Med Systems LLC: Advantage CT/PET Fusion for Advantage Windows Workstation with Volume View CT Colonography/Navigator 2 software. (Version 4.0, 4.1,or 4.2)

Recall number
Z-0854-06
Recalling firm
General Electric Med Systems LLC
Product
Advantage CT/PET Fusion for Advantage Windows Workstation with Volume View CT Colonography/Navigator 2 software. (Version 4.0, 4.1,or 4.2)
Status
Terminated
Initiated
2004-10-08
Posted
2006-05-09
Terminated
2006-12-31
Reason
Display in fused mode- viewer allows 2 studies ( PET/CT) registered in the same spatial domain to be viewed simultaneously. These 2 images should be visually aligned to a precision of less than 1/2 PET Voxel. A problem associated with this version of software may introduce a shift of up to 1.5 PET Voxels, which can be as much as 8 mm with certain zoom factors.
Root cause
Other
Action
An Urgent Safety Notice letter, dated October 8, 2004,(5121309-147, rev.E) was distributed to affected customers and a validated software patch that corrected the problem was installed on affected systems via Field Modification Instruction (FMI) 80114.
Quantity
199
Distribution
Distributed in 32 states: AK, AL, AZ, CA, CO, CT, FL, GA, IL, KY, LA, MA, MD, MI, MN, MO, NC, ND, NH, NJ, NY, OH, OK, PA, SC, TN, TX, UT, VA, WA, WI, WV; Australia, Belgium, Brazil, Canada, Switzerland, Chile, Nigeria, China, France, Egypt, Ireland, Israel, Japan, Korea, the Netherlands, Norway, New Zealand, Phillipines, Sweden, and Taiwan.
Lot, serial or model codes
100943VP0, 67392II1, 36521VE3, 70827VE1, 419453BU6436332BU1, 70800VE8, 44586VE6, 341GE8, 97197VP8, 99813VP8, 97983VP1, 70820VE6, 58626VE3, 70795VE0, FRB3510J25, 58630VE5, 58624VE8, 58589VE3, 7823VE0, 44579VE1, 44597VE3, 44585VE8, 25125VE6, 425022BU1, 70796VE8, 93086VE7, 37621BU4, 76906VE7, 76899VE4, 70798VE4, 65504VE3, 65603VE3, 36543VE7, 58628VE9, 70811VE5, 44584VE1, 44581VE7, FRB3430VBM, 44589VE0, 44598VE1, 44582VE5, 44594VE0, 72740VE4, 1189GE0, 2030GE5, 2085GE0, 2025GE5, 72477II3, 72480II7, 343GE4, 97203VP4, 98272VP8, 423277BU3, 349GE1, 425018BU9, FRB3300M6G, 99699VP1, 98371VP8, 101216VP0, 425015BU5, 237GE8, 364GE0, 339GE2, 101196VP4, 419382BU7, 419383BU5, 366GE5, 367GE3, 357GE4, 2085GE9, 419457BU7, 25128VE0, 98436VP9, 99981VP3, 36513VE0, 98623UP2, 101000VP8, 2028GE9, 2799GE5, 1089GE2, 419450BU2, 101280VP6, 67269VE1, 100855VP6, 66250II2, 346GE7, 363GE2, 345GE9, 1073VP5, 101170VP9, 98516VP8, 2802GE7, 419356BU1, 66244II5, 1487GE8, 67389II7, 365GE7, 425021BU3, 383GE0, 25132VE2, 101334VP1, 95972VP6, 419449BU4, 2031GE3, 2800GE1, 36517VE1, FRB33605L9, 1024GE9, 95307VP5, 102235VP9, 102232VP6, 425204BU5, 1124GE7, 1122GE1, 419455BU1, 419456BU9, 102220VP1, 102555VP0, 97622VP5, 72249II6, 67210VE5, FRB33605KX, 419257BU1, 95268VP9, 338GE4, 101404VP2, 2090GE9, 2087GE5, 101308VP5, 101251VP7, 421989BU5, 342GE6, 99627VP2, 97933VP6, 98653VP9, 66249II4, 72479II9, 25126VE4, 356GE6, 25130VE6, 65450VE9, DE31500056, 347GE5, 421367BU4, 100912VP5, 25123VE1, 25124VE9, 99867VP4, DE24600031, 99910VP2, 101246VP7, 419357BU9, 101171VP7, 102825VP7, 100664VP2, 102856VP2, 424101BU4, 425020BU5, 354GE1, 36518VE9, 100885VP3, 2801GE9, USV41104ST, 2091GE7, USV41103MJ, 99994VP6, 102133VP6, 419549BU1, 25137VE1, 102078VP3, 425351BU4, 1066GE0, 98816VP2, 97591VP2, 99631VP4, 423318BU5, 2086GE7, 97905VP4, 100690VP7, 350GE9, 25139VE7, 1120GE5, 100001VP7, 2084GE2, 391228BU4, 101190VP7, 99909VP4, 1123GE9, 2803GE5, 98474VP0, 25127VE2, 67384II8, 98815VP4, 101288VP9, 101287VP1, 66241II1, 25134VE8
510(k) numbers
K010336
PMA numbers
not on file
Product code
LLZ
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register