Skip to the content

FDA Medical Device Recalls (openFDA)

Beckman Coulter Inc: GEN*S SlideMaker, Barcode Label Specifications, GEN*S SM Intended Use, GEN*S Slide and Sample, Part Numbers 6605360, P/N 4237221, 4237212, and 4277299. Intended for use as an optional peripheral to the COULTER GEN*S System. Creates a blood smear on a clean microscope slide using a segment of the blood sample aspirated by the GEN*S System.

Recall number
Z-0851-2010
Recalling firm
Beckman Coulter Inc
Product
GEN*S SlideMaker, Barcode Label Specifications, GEN*S SM Intended Use, GEN*S Slide and Sample, Part Numbers 6605360, P/N 4237221, 4237212, and 4277299. Intended for use as an optional peripheral to the COULTER GEN*S System. Creates a blood smear on a clean microscope slide using a segment of the blood sample aspirated by the GEN*S System.
Status
Terminated
Initiated
2008-05-29
Posted
2010-02-25
Terminated
2011-11-29
Reason
Beckman Coulter has become aware of a potential issue in which the Sample ID information embedded in the barcode affixed to a slide made by the Coulter GEN*S and LH SlideMaker does not match the text.
Root cause
Pending
Action
An Urgent: Product Corrective Action letter, dated June 9, 2008, was sent to customers. The letter identified the affected products, stated the background, discussed the issue, and stated the actions to be taken. Customers are to: 1. Only use upper case alpha characters in the Sample Identifier 2. If you must use lower case alpha characters in the Sample Identifier, configure the SlideMaker Slide labels without a bar-code. Questions regarding the Product Corrective Action should be directed to Beckman Coulter Customer Service at 1-800-526-7694 or your local Beckman Coulter Rep.
Quantity
4
Distribution
Worldwide Distribution -- USA and Canada.
Lot, serial or model codes
All Software Versions.
510(k) numbers
not on file
PMA numbers
not on file
Product code
KPA
Firm FEI
2050012
Firm city
Brea
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register