Covidien LP: Surgilon Braided Nylon sutures: 88861919-31 SURGILON* 4-0 BLK 7X75CMPCT 88861919-41 SURGILON* 3-0 BLK 7X75CMPCT 88861919-51 SURGILON* 2-0 BLK 7X75CMPCT 88861919-71 SURGILON* 1 BLK 7X75CM PCT
- Recall number
- Z-0844-2024
- Recalling firm
- Covidien LP
- Product
- Surgilon Braided Nylon sutures: 88861919-31 SURGILON* 4-0 BLK 7X75CMPCT 88861919-41 SURGILON* 3-0 BLK 7X75CMPCT 88861919-51 SURGILON* 2-0 BLK 7X75CMPCT 88861919-71 SURGILON* 1 BLK 7X75CM PCT
- Status
- Open, Classified
- Initiated
- 2023-12-21
- Posted
- 2024-01-29
- Terminated
- not on file
- Reason
- Specific lots of sutures were sterilized with gamma doses that exceeded the range approved. The extra gamma levels may decrease the strength of the sutures over time, which could result in harms such as wound dehiscence and hemorrhage/blood loss/ bleeding at a critical level.
- Root cause
- Process control
- Action
- On December 21, the firm notified affected consignees via URGENT MEDICAL DEVICE RECALL letters. Customers were instructed to identify, quarantine, and return all unused, non-expired affected sutures to Medtronic. Please contact rs.covidienfeedbackcustomerservice@medtronic.com for the Return Good Authorization (RGA). Customers will receive credit for returned affected product. If purchased from a distributor, please contact your distributor directly to arrange for the return of product back to your distributor. Pass on the recall notice to all those who need to be aware within your organization, or to any organization where the affected sutures have been transferred or distributed. If you have questions regarding this recall, please call Medtronic Quality Assurance at 800-255-6774 option 1, then option 1 ***Updated 4/1/2024*** Firm identified additional models and lots affected by the gamma radiation issue and sent additional notification to customers on January 25, 2024.
- Quantity
- 46,812 US; 68,976 OUS
- Distribution
- Worldwide US Nationwide distribution including in the states of AL, AZ, CA, FL, GA, HI, IA, IL, IN, MD, MI, MN, MO, MS, NC, NE, NJ, NV, NY, OH, OK, OR, PA, PR, TN, TX, VA, WA, WV and the countries of Australia, Bahrain, Belgium, Bulgaria, Canary Islands, China, Cyprus, Czech Republic, Finland, France, Germany, Greece, Hong Kong, Hungary, Italy, Japan, Korea, Republic Of Kosovo, Mongolia, Netherlands, New Zealand, Panama, Poland, Portugal, Puerto Rico, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, United Kingdom.
- Lot, serial or model codes
- Model 88861919-31 SURGILON* 4-0 BLK 7X75CMPCT UDI-DI 10884521071728, 20884521071725 Lots D1L1929RY D2B2243RY Model 88861919-41 SURGILON* 3-0 BLK 7X75CMPCT UDI-DI 10884521071742, 20884521071749 Lot D2B0176RY Model 88861919-51 SURGILON* 2-0 BLK 7X75CMPCT UDI-DI 10884521071766 Lots D2D2437RY UDI-DI 20884521071763 Lots D2D2437RY D2E0905RY D2E0916RY Model 88861919-71 SURGILON* 1 BLK 7X75CM PCT UDI-DI 20884521071794 Lots D1M0640RY D1M1494RY
- 510(k) numbers
- ["K981582"]
- PMA numbers
- not on file
- Product code
- GAR
- Firm FEI
- 1282497
- Firm city
- Mansfield
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28