Skip to the content

FDA Medical Device Recalls (openFDA)

BIOMERIEUX, INC.: Simplastin HTF Reagents, 20ml HO, ( and 6 ml HO) bioMerieux, Inc., Durham, NC 27704

Recall number
Z-0843-05
Recalling firm
BIOMERIEUX, INC.
Product
Simplastin HTF Reagents, 20ml HO, ( and 6 ml HO) bioMerieux, Inc., Durham, NC 27704
Status
Terminated
Initiated
2005-03-07
Posted
2005-05-18
Terminated
2006-10-16
Reason
The ISI labeling for Simplastin HTF on the Coag-a-Mate MAX and Thrombolyzer instrument platforms is not correct. The Simplastin HTF ISI assignment was mis-assigned and the published values are higher than expected for the listed instrument platforms.
Root cause
Other
Action
Consignees were notified by letter on/about March 7, 2005.
Quantity
19748 kits
Distribution
Nationwide, Canada, Colombia, India, Philippines, Spain, Thailand
Lot, serial or model codes
Product numbers 259846, 259847, Lot numbers: 161849, 161800 and 161798. Recall Expanded (03/23/2005) to include these additional lot numbers: 161655, 161710, 161741, 161738, 161742, 1621739, 161743, 161740, 161764, 161763, 161797 and 161799.
510(k) numbers
K974566
PMA numbers
not on file
Product code
GJS
Firm FEI
3002769706
Firm city
Durham
Firm state
NC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register