Mindray DS USA, Inc. dba Mindray North America: DPM 6 Patient Monitor and DPM 7 Patient Monitor. The product affected is the Mindray IBP module, P/N is 6800-30-50485 for use with Mindray DPM 6 and DPM 7 patient monitors . Intended to be used for monitoring, displaying, reviewing, storing and alarming of multiple physiological parameters.

Recall number
Z-0837-2015
Recalling firm
Mindray DS USA, Inc. dba Mindray North America
Product
DPM 6 Patient Monitor and DPM 7 Patient Monitor. The product affected is the Mindray IBP module, P/N is 6800-30-50485 for use with Mindray DPM 6 and DPM 7 patient monitors . Intended to be used for monitoring, displaying, reviewing, storing and alarming of multiple physiological parameters.
Status
Terminated
Initiated
2014-10-06
Posted
2014-12-19
Terminated
2017-06-06
Reason
Mindray has identified an issue that involves the IBP module used with the DPM 6 and DPM 7 Monitor where the IBP measurement data may be displayed in an incorrect IBP tile on the monitors display.
Root cause
Process control
Action
Mindray DS USA, Inc. sent an Urgent Field Corrective Action Letter dated 10/6/2014 to their customers via certified mail with return receipt. The letter identified the affected product, the problem, adverse effects on the patient, and corrective actions. Customers are asked to either verify that the IBP modules are not affected by following the steps provided in the "Procedures for Verifying IBP Channel Connections;" or to discontinue use of the IBP modules until a Mindray Service Representative contacts them and performs the test to verify the function of the IBP modules. Customers are to contact their local Mindray Service Representative at 1-800-288-2121 to inform them of which corrective action will be taken. Questions should be directed to Diane Arpino at 201-995-8407 or d.arpino@mindray.com.
Quantity
455 units
Distribution
Distributed to the states of AL, CA, CO, FL, GA, IA, IL, IN, LA. MA, MO, MS, MT, NC, ND, NY, OH, PA, SC, TX, VA, WA and WV.
Lot, serial or model codes
DPM 6 - PN: C-047-000199-00 and DPM 7 - PN: C-047-000200-00, IBP Module P/N: 6800-30-50485
510(k) numbers
["K092449"]
PMA numbers
not on file
Product code
MHX
Firm FEI
2221819
Firm city
Mahwah
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register