Skip to the content

FDA Medical Device Recalls (openFDA)

Biocartis Nv: Idylla INSTRUMENT, IVD, REF P0010. Intended for the detection of nucleic acid targets in various types of human samples, using disposable Test-specific ldylla Cartridges.

Recall number
Z-0836-2025
Recalling firm
Biocartis Nv
Product
Idylla INSTRUMENT, IVD, REF P0010. Intended for the detection of nucleic acid targets in various types of human samples, using disposable Test-specific ldylla Cartridges.
Status
Open, Classified
Initiated
2024-12-10
Posted
2025-01-10
Terminated
not on file
Reason
False-positive MSI-H results generated by the Idylla MSI Test.
Root cause
Environmental control
Action
Biocartis issued an URGENT - Medical Device Correction notice to its consignees on 12/10/2024 via email. The notice provided the description of the problem, its impact, and requested the following actions be taken: 1. Until your Instrument is serviced, please move forward with the option that applies to your situation: o In case you have another Idylla Instrument available that is not affected, use that Instrument to run the Idylla MSI Test and discontinue running the Idylla MSI Test on the listed Instrument(s). o No new Idylla MSI testing should be performed using the affected Idylla Instruments. o For results already obtained using an affected Idylla Instrument, the following apply: o MSS results are VALID and may be reported. o MSI-H (Microsatellite Instability High) results with six or seven positive identified biomarkers (as depicted in the result report) are VALID and may be reported. o MSI-H results other than those with six or seven positive identified biomarkers are INVALID and should NOT be reported or used to manage patient health. Patient samples with these results should be re-tested on an Idylla Instrument that is NOT affected. 2. If you are not the Idylla user, please forward this notice to the Idylla end user or Idylla supervisor and provide Biocartis with the correct contact details for our record keeping. 3. Completion of the Acknowledgement of Receipt (Appendix 2) is required for compliance. Please complete and sign the attached Acknowledgement of Receipt form by December 17, 2024, and send it to vigilance@biocartis.com.
Quantity
8 units
Distribution
US: AL, NJ, TX, OH, CA
Lot, serial or model codes
UDI/DI 05415219000119, Serial Numbers: 00002706, 00002820, 00003103, 00003262, 00003419, 00003434, 00003444, and 00003445.
510(k) numbers
not on file
PMA numbers
not on file
Product code
OOI
Firm FEI
3009972873
Firm city
Mechelen
Firm state
not on file
Firm country
Belgium

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register