FDA Medical Device Recalls (openFDA)
Ion Beam Applications S.A.: adaPT insight (12C), v1.3.2, PAT.109 (US), Treatment room 4: I2C is used with a charged particle or photon radiation therapy system for localization of the patient position with respect to the therapy equipment and to provide correction feedback to the radiation therapy device.
- Recall number
- Z-0836-2015
- Recalling firm
- Ion Beam Applications S.A.
- Product
- adaPT insight (12C), v1.3.2, PAT.109 (US), Treatment room 4: I2C is used with a charged particle or photon radiation therapy system for localization of the patient position with respect to the therapy equipment and to provide correction feedback to the radiation therapy device.
- Status
- Terminated
- Initiated
- 2014-11-07
- Posted
- 2014-12-19
- Terminated
- 2015-06-29
- Reason
- Two issues were identified: 1. In the release of adaPT insight v1.3.2., the kV/kV Single Source acquisition workflow was linked to a wrong geometrical calibration file, resulting in an alignment offset for this specific mode of acquisition. 2. A gantry collision incident required the IBA service team to verify the gantry alignment and protons I x-ray beams colinearity, du
- Root cause
- Software design (manufacturing process)
- Action
- The firm, IBA, sent a " User Notice" dated November 7, 2014, to the sole affected site-UPHS Customer. The notice describes the product, problem and actions to be taken. The customer was instructed that the KV/kV source acquisition mode has been disabled and shall not be used for patient alignment until system is officially confirmed to be back within tolerances. IBA will change the kV/kV Single Source acquisition workflow configuration file to be linked to the correct geometrical calibration file and will update the kV/kV Single Source calibration file that includes the colinearity parameters, with the correct values. If you need assistance, contact the IBA Site Staff or the Operations manager at +32 10 475 997 or email: Sophie.dessart@iba-group.com.
- Quantity
- 1
- Distribution
- US Distribution in PA only.
- Lot, serial or model codes
- PAT.109 (US), Treatment room 4
- 510(k) numbers
- ["K132847"]
- PMA numbers
- not on file
- Product code
- LHN
- Firm FEI
- 3000256071
- Firm city
- Louvain La Neuve
- Firm state
- not on file
- Firm country
- Belgium
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28