Philips Medical Systems (Cleveland) Inc: Brilliance iCT & Brilliance iCT SP Computed Tomography X-Ray Systems, Philips Healthcare, Cleveland, OH The Brilliance iCT and iCT SP are whole body Computed Tomography X-Ray Systems intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient and equipment supports, components and accessories.
- Recall number
- Z-0832-2014
- Recalling firm
- Philips Medical Systems (Cleveland) Inc
- Product
- Brilliance iCT & Brilliance iCT SP Computed Tomography X-Ray Systems, Philips Healthcare, Cleveland, OH The Brilliance iCT and iCT SP are whole body Computed Tomography X-Ray Systems intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient and equipment supports, components and accessories.
- Status
- Terminated
- Initiated
- 2013-12-09
- Posted
- 2014-01-27
- Terminated
- 2014-05-29
- Reason
- Patient images exhibited ring artifacts.
- Root cause
- Software design
- Action
- The firm, Philips Healthcare, sent an "URGENT - Medical Device Correction" letter dated December 9, 2013 to their customers. The letter described the product, problem and actions to be taken. The customers were instructed to do the following to avoid the occurrence of these specific ring artifacts: 1. Disable Automatic Scan Time for all Exam Cards 2. Leave Automatic Scan Time on during planning and prior to clicking Go to initiate the scan, click the acquisition series of the scan and disable Automatic Scan Time, then proceed with the scan. 3. Disable Edit Before Final Recon (Reconstruction) for all Exam Cards If a ring artifact is present, clinical judgment should be used by the customer/user to determine if the images acquired in the dataset contain enough information to make a diagnosis. Philips will be releasing a Field Change Order (FCO) to correct the issue with software version release 4.1.1.19261. In addition, a Philips Field Service Engineer will contact you to schedule the software installation at your site. If you need any further information or support concerning this issue, please contact your local Philips representative or local Philips Healthcare office. For North America and Canada, contact the Customer Care Solutions Center (1-800-722-9377, Option 5: Enter Site ID or follow the prompts).
- Quantity
- 7 units
- Distribution
- Worldwide Distribution: US (nationwide) including states of: FL, IN, NY, OR and PA; and internationally to: Australia and Israel.
- Lot, serial or model codes
- Brilliance iCT: 728306, Serial #'s: 100018, 100040, 100411, 100503, 100506 Brilliance iCT SP: 728311, Serial #'s: 200047, 200121
- 510(k) numbers
- ["K060937"]
- PMA numbers
- not on file
- Product code
- JAK
- Firm FEI
- 1525965
- Firm city
- Cleveland
- Firm state
- OH
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28