Philips Medical Systems (Cleveland) Inc: Brilliance iCT & Brilliance iCT SP Computed Tomography X-Ray Systems, Philips Healthcare, Cleveland, OH The Brilliance iCT and iCT SP are whole body Computed Tomography X-Ray Systems intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient and equipment supports, components and accessories.

Recall number
Z-0832-2014
Recalling firm
Philips Medical Systems (Cleveland) Inc
Product
Brilliance iCT & Brilliance iCT SP Computed Tomography X-Ray Systems, Philips Healthcare, Cleveland, OH The Brilliance iCT and iCT SP are whole body Computed Tomography X-Ray Systems intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient and equipment supports, components and accessories.
Status
Terminated
Initiated
2013-12-09
Posted
2014-01-27
Terminated
2014-05-29
Reason
Patient images exhibited ring artifacts.
Root cause
Software design
Action
The firm, Philips Healthcare, sent an "URGENT - Medical Device Correction" letter dated December 9, 2013 to their customers. The letter described the product, problem and actions to be taken. The customers were instructed to do the following to avoid the occurrence of these specific ring artifacts: 1. Disable Automatic Scan Time for all Exam Cards 2. Leave Automatic Scan Time on during planning and prior to clicking Go to initiate the scan, click the acquisition series of the scan and disable Automatic Scan Time, then proceed with the scan. 3. Disable Edit Before Final Recon (Reconstruction) for all Exam Cards If a ring artifact is present, clinical judgment should be used by the customer/user to determine if the images acquired in the dataset contain enough information to make a diagnosis. Philips will be releasing a Field Change Order (FCO) to correct the issue with software version release 4.1.1.19261. In addition, a Philips Field Service Engineer will contact you to schedule the software installation at your site. If you need any further information or support concerning this issue, please contact your local Philips representative or local Philips Healthcare office. For North America and Canada, contact the Customer Care Solutions Center (1-800-722-9377, Option 5: Enter Site ID or follow the prompts).
Quantity
7 units
Distribution
Worldwide Distribution: US (nationwide) including states of: FL, IN, NY, OR and PA; and internationally to: Australia and Israel.
Lot, serial or model codes
Brilliance iCT: 728306, Serial #'s: 100018, 100040, 100411, 100503, 100506  Brilliance iCT SP: 728311, Serial #'s: 200047, 200121
510(k) numbers
["K060937"]
PMA numbers
not on file
Product code
JAK
Firm FEI
1525965
Firm city
Cleveland
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register