FDA Medical Device Recalls (openFDA)
Medtronic: Covidien Endo GIA Ultra Universal Standard Stapler Handle Ref: EGIAUSTND Product Usage: The Endo GIA Ultra Universal Short, Endo GIA Ultra Universal and Endo GIA Ultra Universal XL staplers have applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection and creation of anastomosis. It may be used for transection and resection of liver substance, hepatic vasculature and biliary structures. The Endo GIA Ultra Universal Short, Endo GIA Ultra Universal and Endo GIA Ultra Universal XL staplers when used with the Endo GIA curved tip single use reloads can be used to blunt dissect or separate target tissue from other tissue.
- Recall number
- Z-0820-2016
- Recalling firm
- Medtronic
- Product
- Covidien Endo GIA Ultra Universal Standard Stapler Handle Ref: EGIAUSTND Product Usage: The Endo GIA Ultra Universal Short, Endo GIA Ultra Universal and Endo GIA Ultra Universal XL staplers have applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection and creation of anastomosis. It may be used for transection and resection of liver substance, hepatic vasculature and biliary structures. The Endo GIA Ultra Universal Short, Endo GIA Ultra Universal and Endo GIA Ultra Universal XL staplers when used with the Endo GIA curved tip single use reloads can be used to blunt dissect or separate target tissue from other tissue.
- Status
- Terminated
- Initiated
- 2015-12-02
- Posted
- 2016-02-18
- Terminated
- 2019-07-18
- Reason
- Staplers fail to fire or partially fire and reports of the instrument articulating lever disengaging during use
- Root cause
- Process control
- Action
- Medtronic sent an Urgent Medical Device Recall letters dated December 2 2015 and an Updated Urgent Medical Device Recall letter dated January 13, 2016, to affected customers, via Federal Express. The letter identified the affected product, problem and actions to be taken. Customers were instructed to discontinue use of the product and return all inventory to Medtronic (Covidien), 195 McDermott Road, North Haven, CT 06473. For questions or concerns, please do not hesitate to contact your Medtronic representative or Customer Service at (800) 882-5878.
- Quantity
- 1,102,166 units
- Distribution
- Worldwide Distribution - US Nationwide and the countries of Canada, Albania, ALGERIA, Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Chile, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Gabon, Georgia, Germany, Greece, Greece, Guatemala, Honduras, Hungary, Iceland, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Mauritius, Mexico, Monaco, Montenegro, Morocco, Namibia, Netherlands, Nicaragua, Nigeria, Norway, Panama, Paraguay, Peru, Poland, Portugal, Romania, Russia, Russian Federation, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Tunisia, Turkey, UAE, Ukraine, United Kingdom, Uruguay, Venezuela, and Zimbabwe.
- Lot, serial or model codes
- RANGES OF LOT NUMBERS (all lot numbers within each range are affected) Lot # starting with P3C through P5B0521X P5D0410X through P5D0411X P5D0823X through P5D0829X P5E0080X through P5E0082X P5E0317X through P5E0320X P5E0322X through P5E0324X P5F0001X through P5F0002X INDIVIDUAL LOT NUMBERS (in addition to ranges) P5C0239X, P5C0241X, P5C0508X, P5C0510X, P5C0512X, P5C0597X P5C0599X, P5C0601X, P5C0603X, P5C0684X, P5C0803X, P5C0805X, P5C0807X. P5E0527X, P5E0529X P5D0067X, P5D0069X , P5D0073X, P5D0170X, P5D0413X, P5D0415X ,P5D0417X, P5D0419X, P5D0447X, P5D0623X, P5D0624X ,P5D0626X, P5D0628X P5D0630X, P5E0525X, P5F0004X, P5F0006X, P5F0008X, P5F0010X, P5F0014X, P5F0017X, P5F0019X, P5F0021X, P5F0213X, P5F0217X
- 510(k) numbers
- K111825
- PMA numbers
- not on file
- Product code
- GDW
- Firm FEI
- 1219930
- Firm city
- North Haven
- Firm state
- CT
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- 3e6df46828c1
- Size
- 288213574 bytes
- Published
- 2026-10-06
- Read
- 2026-10-07T03:20:25.099Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-14
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 3e6df46828c1, published 2026-10-06