FDA Medical Device Recalls (openFDA)
NAImco Inc dba Richmar Inc: Cloth electrodes with Carbon construction: P/N 201-1231, Cat Nos.: 201-123, 202-454, 203-118, 203-342, 203-454, 203-493, 203-672, 400-877, and 400-877-BJC, 2" x 2" Electrode, Cloth, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1241, Cat Nos.: 201-124, 202-455, 203-340, 203-674, and 400-872, 2" Round Electrode, Cloth, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1251, Cat Nos.: 201-138, 202-276, 202-456, 203-336, 203-680, and 400-879, 3" Round Electrode, Cloth, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1381, Cat Nos.: 201-126, 203-678, and 400-856, 2" x 3.5" Electrode, Cloth, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1501, Cat Nos.: 201-150 and 400-863, 1.25" Round Electrode, Cloth, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1521, Cat No. 400-852, 3 x 5 Electrode, Cloth, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1541, Cat Nos.:201-154 and 400-865 1.5 x 2.5 Oval Electrode, Cloth, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 Cutaneous electrode.
- Recall number
- Z-0815-2013
- Recalling firm
- NAImco Inc dba Richmar Inc
- Product
- Cloth electrodes with Carbon construction: P/N 201-1231, Cat Nos.: 201-123, 202-454, 203-118, 203-342, 203-454, 203-493, 203-672, 400-877, and 400-877-BJC, 2" x 2" Electrode, Cloth, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1241, Cat Nos.: 201-124, 202-455, 203-340, 203-674, and 400-872, 2" Round Electrode, Cloth, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1251, Cat Nos.: 201-138, 202-276, 202-456, 203-336, 203-680, and 400-879, 3" Round Electrode, Cloth, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1381, Cat Nos.: 201-126, 203-678, and 400-856, 2" x 3.5" Electrode, Cloth, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1501, Cat Nos.: 201-150 and 400-863, 1.25" Round Electrode, Cloth, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1521, Cat No. 400-852, 3 x 5 Electrode, Cloth, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1541, Cat Nos.:201-154 and 400-865 1.5 x 2.5 Oval Electrode, Cloth, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 Cutaneous electrode.
- Status
- Terminated
- Initiated
- 2009-11-30
- Posted
- 2013-02-14
- Terminated
- 2013-09-03
- Reason
- Electrode Socket Terminal pulling away from the electrode lead wire when the electrical stim lead wire was removed.
- Root cause
- Unknown/Undetermined by firm
- Action
- Rich-Mar sent "URGENT: MEDICAL DEVICE VOLUNTARY RECALL" letters to affected customers beginning on 11/30/2009. The letter identified the product, problem, and actions to be taken by the customers. Contact the firm at 423-648-7730 ext. 214 for questions relating to t his recall.
- Quantity
- 209860 electrodes
- Distribution
- Distributed in the states of GA and TN.
- Lot, serial or model codes
- All Serial Numbers distributed between 11/01/2009 and 01/31/2010
- 510(k) numbers
- ["K050469"]
- PMA numbers
- not on file
- Product code
- GXY
- Firm FEI
- 3004750362
- Firm city
- Chattanooga
- Firm state
- TN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28