Skip to the content

FDA Medical Device Recalls (openFDA)

AGFA Healthcare Corp.: Radiomat M+ NIF 14 x 17, Radiomat M+ NIF 35CM x 43CM, Radiomat M+ NIF 10 x 12, Radiomat M+ NIF 25,4CM x 30,5CM

Recall number
Z-0804-2015
Recalling firm
AGFA Healthcare Corp.
Product
Radiomat M+ NIF 14 x 17, Radiomat M+ NIF 35CM x 43CM, Radiomat M+ NIF 10 x 12, Radiomat M+ NIF 25,4CM x 30,5CM
Status
Terminated
Initiated
2014-10-24
Posted
2014-12-16
Terminated
2015-03-16
Reason
Some material of the coating Type CXCPMV3 has potential artifacts showing a small bright thin line when the film is processed.
Root cause
Other
Action
Dealers were contacted by Agfa HealthCare via phone and/or email on 10/24/2014 and the week of 11/10/14 to inform them of the quality issue and to quarantine identified product shipped to their facility or end users. On 11/17/2014, an "URGENT SAFETY NOTICE" letter was sent to the US dealers. The letter described the safety alert and the request to quarantine affected product. Acknowledgment via FAX-Back or email that the information was received and understood was requested from the dealers. A customer Urgent Field Safety Notice was also provided to those dealers who had shipped affected product to end users so they could sent it to the end use customers.
Quantity
142 units
Distribution
US Nationwide Distribution in the states of: AL, AZ, CA, LA, MI, NE, NH, NJ, NC, VA
Lot, serial or model codes
Lot 79440009
510(k) numbers
not on file
PMA numbers
not on file
Product code
IWZ
Firm FEI
3001236302
Firm city
Greenville
Firm state
SC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-14

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this source