Skip to the content

FDA Medical Device Recalls (openFDA)

Nellcor Puritan Bennett Inc. (dba Covidien LP): BIS Bilateral Sensors, part number 186-0212, packaged in boxes of 10. For use with BIS Monitoring System, Labeled as either Covidien Llc, or Aspect Medical Systems, Inc. Product Usage: To enable recordings of electrophysiological (such as EEG) signals. The Sensor is a single patient use, disposable, pre-gelled 6 electrode array that is applied directly to the patients forehead. BIS technology is used as an adjunct to clinical judgement and training. Reliance on BIS technology alone for intraoperative anesthetic management is not recommended.

Recall number
Z-0802-2012
Recalling firm
Nellcor Puritan Bennett Inc. (dba Covidien LP)
Product
BIS Bilateral Sensors, part number 186-0212, packaged in boxes of 10. For use with BIS Monitoring System, Labeled as either Covidien Llc, or Aspect Medical Systems, Inc. Product Usage: To enable recordings of electrophysiological (such as EEG) signals. The Sensor is a single patient use, disposable, pre-gelled 6 electrode array that is applied directly to the patients forehead. BIS technology is used as an adjunct to clinical judgement and training. Reliance on BIS technology alone for intraoperative anesthetic management is not recommended.
Status
Terminated
Initiated
2011-11-08
Posted
2012-01-26
Terminated
2013-03-13
Reason
During manufacture, circuit traces were reversed which causes inaccurate display of information.
Root cause
Finished device change control
Action
Covidien sent an Urgent Medical Device Recall letter dated November 8, 2011, to all affected customers. The letter identified the product, problem, and actions to be taken. Customers were instructed to segregate and return any product with the affected lot numbers according to the directions provided. If product was further distributed, the recall was to be continued to those customers. Further information is available at 800-635-5267.
Quantity
8510 units (851 boxes of 10)
Distribution
Nationwide Distribution (USA) including the states of: AZ, CA, CO, FL, GA, IL, KY, KS, LA, MA, MO, MS, NC, NM, NY, OH, PA, SD, TN, TX, UT, VA, WA, WI, WV and WY.
Lot, serial or model codes
Lot numbers: 0528111A, 0603111A, 0604111A, 0606111A, 0607111A, 0608111A, 0609111A, 0615111A, 0620111A, 0621111A, 0624111A, 0625111A, 0627111A, 0701111A, 0707111A, 0712111A, 0713111A, 0725111A, 0726111A, 0726111B, 0727111A, 0727111B, 0728111A, 0728111B, 0729111A, 0729111B, 0801111A, 0808111A, 0809111A, 0810111A, 0811111A, 0812111A, 0813111A, 0814111A, 0820111A, 0821111A, 0822111A, 0822111B, 0823111A, 0823111B, 0824111A, 0824111B, 0825111A, 0825111B, 0826111A, 0826111B, 0921111A, 0923111A, 0923111B, 0924111A, 0926111A, 0927111A, 0930111A, 1001111A, 1003111A, 1004111A, 1005111A, 1006111A.
510(k) numbers
K062692
PMA numbers
not on file
Product code
GXY
Firm FEI
2936999
Firm city
Boulder
Firm state
CO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register