Skip to the content

FDA Medical Device Recalls (openFDA)

Abbott Laboratories MPG: ARCHITECT Folate Reagent Kit, list 6C12-20 ( 4 x 100 tests), 6C12-25 (1 x 100 tests) and 6C12-30 (4 x 500 tests), Abbott Laboratories, Abbott Park, IL 600064 U.S.A.

Recall number
Z-0783-05
Recalling firm
Abbott Laboratories MPG
Product
ARCHITECT Folate Reagent Kit, list 6C12-20 ( 4 x 100 tests), 6C12-25 (1 x 100 tests) and 6C12-30 (4 x 500 tests), Abbott Laboratories, Abbott Park, IL 600064 U.S.A.
Status
Terminated
Initiated
2005-03-11
Posted
2005-04-28
Terminated
2006-01-18
Reason
concentrations for normal samples generated with current reagent lots of the ARCHITECT Folate assay are lower than data generated for the package insert.
Root cause
Other
Action
Abbott international affilates were notified of the recall by e-mail on 03/11/2005, and requested to notifiy their customers. The letters inform the end users of the downward shift in control and/or patient results, and requested them to assess their normal and deficient ranges for their specific population, assess their quality control ranges and follow their laboratory procedures for reviewing patient results and notifying Health Care Providers that they serve. There was no U.S. distribution.
Quantity
2,521 kits
Distribution
Canada, Singapore, New Zealand, Germany, Hong Kong, Japan, and Australia
Lot, serial or model codes
lot numbers 21082M200, 22024M100, 21082M201, 22024M101, 16603M200, and 22025M100
510(k) numbers
not on file
PMA numbers
not on file
Product code
CGN
Firm FEI
1415939
Firm city
Abbott Park
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register