Skip to the content

FDA Medical Device Recalls (openFDA)

MEDLINE INDUSTRIES, LP - Northfield: Medline custom medical procedure kits labeled as: 1) CATH LAB PACK-LF, REF DYNJ0382559M; 2) CRANI, REF DYNJ56001B; 3) GENDER AFFIRMING SURGERY, REF DYNJ66252; 4) GENDER AFFIRMING SURGERY, REF DYNJ66252A; 5) MAJOR NEURO PACK-LF, REF DYNJ0578916X.

Recall number
Z-0777-2025
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Product
Medline custom medical procedure kits labeled as: 1) CATH LAB PACK-LF, REF DYNJ0382559M; 2) CRANI, REF DYNJ56001B; 3) GENDER AFFIRMING SURGERY, REF DYNJ66252; 4) GENDER AFFIRMING SURGERY, REF DYNJ66252A; 5) MAJOR NEURO PACK-LF, REF DYNJ0578916X.
Status
Open, Classified
Initiated
2024-11-21
Posted
2024-12-30
Terminated
not on file
Reason
Custom convenience kits are being recall because the contain non-sterile Codman Surgical Patties & Surgical Strips, which have the potential for endotoxin contamination.
Root cause
Nonconforming Material/Component
Action
Medline issued a MEDICAL DEVICE VOLUNTARY RECALL notice on 11/21/2024 to its consignees via US Mail and email. The notice explained the issue, potential risk to health, and requested that the affected product be quarantined. Once the response is received, Medline will provide labels to place on affected inventory, with instructions for the consignee to remove the affected component prior to using the kit.
Quantity
1236 units
Distribution
Worldwide distribution - US Nationwide and the countries of UAE, and Canada.
Lot, serial or model codes
1) REF DYNJ0382559M, UDI/DI 10889942063994 (EA) 10889942063995 (CS), Lot Numbers: 19TBC792; 2) REF DYNJ56001B, UDI/DI 10889942798278 (EA) 40889942798279 (CS), Lot Numbers: 19BDC737, 19FDA310, 19FDB230, 19JDA441; 3) REF DYNJ66252, UDI/DI 10193489324785 (EA) 40193489324786 (CS), Lot numbers: 20DBK065, 20FBN251, 20HBL176, 20HBM692, 21CBP776; 4) REF DYNJ66252A, UDI/DI 10193489887853 (EA) 40193489887854 (CS), Lot numbers: 21IBN151, 21KBO915, 21LBJ105, 21VBB714; 5) REF DYNJ0578916X, UDI/DI 10193489288384 (EA) 40193489288385 (CS), Lot numbers: 22OBL262,20CKA219, 20CKA875, 20EKB076, 20GKA543, 20GKB272, 20IKB497, 20JKB716, 21AKA500, 21BKA195, 21FMD808, 21HMD657, 21IMB172, 21JMH627, 21KMH877, 21LMC161, 22AMA533, 22DMC612, 22DMI233, 22GMB685, 22IMA041, 23AMD730, 23DMA094, 23FMF542, 23IME618, 24AMA230.
510(k) numbers
not on file
PMA numbers
not on file
Product code
OES
Firm FEI
1417592
Firm city
Northfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register