Deerfield Imaging, Inc.: iMRX MRI and Angiography System, part number 10094137 and 10094141, a surgical/interventional suite with integrated MRI and X-ray angiography. The IMRIS iMRX system is the integration of a Siemens Artis zee family (biplane or floor or ceiling mounted single plane) and the IMRIS Neuro 3T System. The System is a solution providing MRI scanning capabilities in the angiography room while retaining all functionality of Siemens Artis zee system. The System includes a traditional MRI unit that has been suspended on an overhead rail system, and is designed to operate inside RF room to facilitate intra-operative, interventional, and multi-room imaging. It also allows intra-operative or interventional imaging capabilities, while retaining all diagnostic functionality of a standard MRI system in the diagnostic room. The iMRX system allows for the sequential use of either MR imaging or X-ray fluoroscopic imaging of the head or chest for a single patient while positioned on a common fixed table either prior to, during or post clinical procedure. When these images are interpreted by a trained physician, they may assist in diagnosis during either neurovascular or cardiovascular interventionsa procedures. The iMRX system is designed such that the imaging sub-systems cannot operate simultaneously and only in accordance with their respective cleared intended use.

Recall number
Z-0762-2020
Recalling firm
Deerfield Imaging, Inc.
Product
iMRX MRI and Angiography System, part number 10094137 and 10094141, a surgical/interventional suite with integrated MRI and X-ray angiography. The IMRIS iMRX system is the integration of a Siemens Artis zee family (biplane or floor or ceiling mounted single plane) and the IMRIS Neuro 3T System. The System is a solution providing MRI scanning capabilities in the angiography room while retaining all functionality of Siemens Artis zee system. The System includes a traditional MRI unit that has been suspended on an overhead rail system, and is designed to operate inside RF room to facilitate intra-operative, interventional, and multi-room imaging. It also allows intra-operative or interventional imaging capabilities, while retaining all diagnostic functionality of a standard MRI system in the diagnostic room. The iMRX system allows for the sequential use of either MR imaging or X-ray fluoroscopic imaging of the head or chest for a single patient while positioned on a common fixed table either prior to, during or post clinical procedure. When these images are interpreted by a trained physician, they may assist in diagnosis during either neurovascular or cardiovascular interventionsa procedures. The iMRX system is designed such that the imaging sub-systems cannot operate simultaneously and only in accordance with their respective cleared intended use.
Status
Terminated
Initiated
2019-11-26
Posted
not on file
Terminated
2020-08-04
Reason
There is a potential hardware error for the iMRX System with a generator A100. A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible
Root cause
Component design/selection
Action
An Urgent Field Safety Notice letter was sent to customers on 11/26/2019 via FedEx.. The letter state the issue, potential risk, the actions to be taken, and stated that the device will be modified. Questions can be directed to the service line at 1-866-475-0525.
Quantity
6
Distribution
CT, MA, VA. International distribution: Australia, Canada, and Qatar.
Lot, serial or model codes
Serial Number: 146792, 153374, 153642, 153793, 153896, 154588.
510(k) numbers
K101813
PMA numbers
not on file
Product code
LNH
Firm FEI
3010326005
Firm city
Minnetonka
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register