FDA Medical Device Recalls (openFDA)
Del Mar Reynolds Medical, Ltd.: Affected device is Model 98400-SL6-AHA, CardioExpress Electrocardiograph. The ECG cable part number is 01.57.471017. The electrocardiograph is only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The intended use of electrocardiograph is to acquire ECG signals from adult and pediatric patients through body surface with ECG electrodes. The electrocardiograph is only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by the electrocardiograph can help users to analyze and diagnose heart disease. However, the interpreted ECG with measurements and interpretive statements is only intended to be used on adult patients and is offered to clinicians on an advisory basis only.
- Recall number
- Z-0743-2012
- Recalling firm
- Del Mar Reynolds Medical, Ltd.
- Product
- Affected device is Model 98400-SL6-AHA, CardioExpress Electrocardiograph. The ECG cable part number is 01.57.471017. The electrocardiograph is only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The intended use of electrocardiograph is to acquire ECG signals from adult and pediatric patients through body surface with ECG electrodes. The electrocardiograph is only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by the electrocardiograph can help users to analyze and diagnose heart disease. However, the interpreted ECG with measurements and interpretive statements is only intended to be used on adult patients and is offered to clinicians on an advisory basis only.
- Status
- Terminated
- Initiated
- 2010-09-10
- Posted
- 2012-01-13
- Terminated
- 2012-01-13
- Reason
- CardioExpress SL6 ECG cable provided with the electrocardiograph is mislabeled. The left arm (LA) is mislabeled as right leg (RL). The left leg (LL) is mislabeled as left arm (LA). The right leg (RL) is mislabeled as left leg (LL).
- Root cause
- Employee error
- Action
- Spacelabs Healthcare sent an "URGENT-MEDICAL DEVICE CORRECTION" letter dated September 24, 2010 to one affected customer in Massachusetts. The letter described the product, problem, and actions to be taken by the customers. Customers were instructed to review previous diagnoses made from this electrocardiograph to mitigate any inappropriate recordings or treatment made, and dispose of the mislabeled ECG cable to prevent its use. Spacelabs replaced the affected ECG cable with the new ECG cable. If you have any questions about this corrective action program, please contact Spacelabs Healthcare at 1-800-522-7025 and select 2 for Technical Support.
- Quantity
- 15 units
- Distribution
- Nationwide distribution.
- Lot, serial or model codes
- Serial numbers: 98400-SL6-AHA1030508, 98400-SL6-AHA1030509, 98400-SL6-AHA1030510, 98400-SL6-AHA1030511, 98400-SL6-AHA1030512, 98400-SL6-AHA1030513, 98400-SL6-AHA1030514, 98400-SL6-AHA1030515, 98400-SL6-AHA1030516, 98400-SL6-AHA1030517, 98400-SL6-AHA1030518, 98400-SL6-AHA1030519, 98400-SL6-AHA1030520, 98400-SL6-AHA1030521, 98400-SL6-AHA1030522.
- 510(k) numbers
- ["K090367"]
- PMA numbers
- not on file
- Product code
- DPS
- Firm FEI
- 3002807992
- Firm city
- Hertford
- Firm state
- not on file
- Firm country
- United Kingdom
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28