FDA Medical Device Recalls (openFDA)

Del Mar Reynolds Medical, Ltd.: Affected device is Model 98400-SL6-AHA, CardioExpress Electrocardiograph. The ECG cable part number is 01.57.471017. The electrocardiograph is only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The intended use of electrocardiograph is to acquire ECG signals from adult and pediatric patients through body surface with ECG electrodes. The electrocardiograph is only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by the electrocardiograph can help users to analyze and diagnose heart disease. However, the interpreted ECG with measurements and interpretive statements is only intended to be used on adult patients and is offered to clinicians on an advisory basis only.

Recall number
Z-0743-2012
Recalling firm
Del Mar Reynolds Medical, Ltd.
Product
Affected device is Model 98400-SL6-AHA, CardioExpress Electrocardiograph. The ECG cable part number is 01.57.471017. The electrocardiograph is only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The intended use of electrocardiograph is to acquire ECG signals from adult and pediatric patients through body surface with ECG electrodes. The electrocardiograph is only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by the electrocardiograph can help users to analyze and diagnose heart disease. However, the interpreted ECG with measurements and interpretive statements is only intended to be used on adult patients and is offered to clinicians on an advisory basis only.
Status
Terminated
Initiated
2010-09-10
Posted
2012-01-13
Terminated
2012-01-13
Reason
CardioExpress SL6 ECG cable provided with the electrocardiograph is mislabeled. The left arm (LA) is mislabeled as right leg (RL). The left leg (LL) is mislabeled as left arm (LA). The right leg (RL) is mislabeled as left leg (LL).
Root cause
Employee error
Action
Spacelabs Healthcare sent an "URGENT-MEDICAL DEVICE CORRECTION" letter dated September 24, 2010 to one affected customer in Massachusetts. The letter described the product, problem, and actions to be taken by the customers. Customers were instructed to review previous diagnoses made from this electrocardiograph to mitigate any inappropriate recordings or treatment made, and dispose of the mislabeled ECG cable to prevent its use. Spacelabs replaced the affected ECG cable with the new ECG cable. If you have any questions about this corrective action program, please contact Spacelabs Healthcare at 1-800-522-7025 and select 2 for Technical Support.
Quantity
15 units
Distribution
Nationwide distribution.
Lot, serial or model codes
Serial numbers:  98400-SL6-AHA1030508, 98400-SL6-AHA1030509, 98400-SL6-AHA1030510, 98400-SL6-AHA1030511, 98400-SL6-AHA1030512, 98400-SL6-AHA1030513, 98400-SL6-AHA1030514, 98400-SL6-AHA1030515, 98400-SL6-AHA1030516, 98400-SL6-AHA1030517, 98400-SL6-AHA1030518, 98400-SL6-AHA1030519, 98400-SL6-AHA1030520, 98400-SL6-AHA1030521, 98400-SL6-AHA1030522.
510(k) numbers
["K090367"]
PMA numbers
not on file
Product code
DPS
Firm FEI
3002807992
Firm city
Hertford
Firm state
not on file
Firm country
United Kingdom

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register