Skip to the content

FDA Medical Device Recalls (openFDA)

Asahi Medical Co Ltd: Asahi APS Series Dialyzers, Model Nos. APS-400S, APS-550S, APS-650S, APS-750S, APS-900S, APS-1050S; the dialyzers are intended for single or initial use and under reprocessing and reuse conditions fo rup to 15 reuse cycles; Manufacturer: Asahi Medical Co., Ltd., 9-1, Kanda Mitoshirocho, Chiyoda-ku, Tokyo 101-8482, Japan; Distributed in teh USA by Nextron Medical, 45 Kulick Road, Fairfield, NJ 07004 USA

Recall number
Z-0739-04
Recalling firm
Asahi Medical Co Ltd
Product
Asahi APS Series Dialyzers, Model Nos. APS-400S, APS-550S, APS-650S, APS-750S, APS-900S, APS-1050S; the dialyzers are intended for single or initial use and under reprocessing and reuse conditions fo rup to 15 reuse cycles; Manufacturer: Asahi Medical Co., Ltd., 9-1, Kanda Mitoshirocho, Chiyoda-ku, Tokyo 101-8482, Japan; Distributed in teh USA by Nextron Medical, 45 Kulick Road, Fairfield, NJ 07004 USA
Status
Terminated
Initiated
2001-10-01
Posted
2004-07-20
Terminated
2004-03-24
Reason
Customer reprocessing methods for the reuseable hollow fiber dialyzers may damage the hollow fibers and result in blood leaks during dialysis.
Root cause
Other
Action
Asahi visited each of the U.S. customers starting on 10/1/01, observing how the facility handles the dialyzer before pre-cleaning, during pre-cleaning and priming. The users were retrained in the reprocessing steps, emphasizing the controls and care to be taken during pre-cleaning, including the verification of water pressure.
Quantity
180,000 units estimated
Distribution
New Jersey, Ohio, Texas, California, Georgia, Pennsylvania, Illinois and Indiana
Lot, serial or model codes
Model Nos. APS-400S, APS-550S, APS-650S, APS-750S, APS-900S, APS-1050S. The recall is no lot specific.
510(k) numbers
["K001250"]
PMA numbers
not on file
Product code
KDI
Firm FEI
3002806407
Firm city
Chiyoda Ku
Firm state
not on file
Firm country
Japan

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register