Skip to the content

FDA Medical Device Recalls (openFDA)

INTELERAD MEDICAL SYSTEMS INCORPORATED: IntelePACS - InteleConnect / TechPortal

Recall number
Z-0723-2026
Recalling firm
INTELERAD MEDICAL SYSTEMS INCORPORATED
Product
IntelePACS - InteleConnect / TechPortal
Status
Open, Classified
Initiated
2025-10-17
Posted
2025-11-26
Terminated
not on file
Reason
Software application that receives digital images and data to be communicated, processed, manipulated, enhanced, stored, displayed has a bug that could cause data loss in Relevant Clinical Info field when modifying studies through Case Editor in either Technologist Portal or Referring Physician Portal, which could result in loss of clinical information, which could impact clinical decision-making.
Root cause
Software Design Change
Action
On 10/17/2025, correction notices were emailed to customers who were asked to do the following: 1. Schedule Feature Release Upgrade (FRU) update immediately with the firm's Customer Success Manager (CSM). 2. Interim Mitigation: Until the update is applied, please advise all staff to verify the 'Relevant Clinical Info' field after any Case Editor modifications and consider temporarily documenting critical clinical information in alternative fields as backup. 3. complete and return the Acknowledgement Form via email to regulatory-affairs@intelerad.com. For more information or to ask any questions about the notification contact the firm's regulatory affairs department at regulatory-affairs@intelerad.com
Quantity
19
Distribution
US Nationwide distribution in the states of AZ, CT, AL, IL, WA, MD, NJ,CA, TN, IN, FL, NC.
Lot, serial or model codes
UDI-DI: B228INTELEPACS0. Version/Dates of Distribution: PACS-5-6-1-R64 through R68/02 Jul 2025 - 10 Sep 2025, PACS-5-7-1-R42 through R47/26 June 2025 - 25 Sep 2025, PACS 5-8-1-R20 through R23/26 Jun 2025 - 25 Sep 2025, PACS-5-9-1-R1 through R3/30 Jun 2025 - 09 Sep 2025
510(k) numbers
["K192176"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
3003768392
Firm city
Montreal
Firm state
not on file
Firm country
Canada

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register