Beckman Coulter Inc.: MicroScan Pos MIC Panel Type 34 - Product Usage: For the determination of antimicrobial agent susceptibility and/or identification to the species level of rapidly growing aerobic and facultative gram-positive cocci, some fastidious aerobic cocci and listeria monocytogenes.

Recall number
Z-0720-2021
Recalling firm
Beckman Coulter Inc.
Product
MicroScan Pos MIC Panel Type 34 - Product Usage: For the determination of antimicrobial agent susceptibility and/or identification to the species level of rapidly growing aerobic and facultative gram-positive cocci, some fastidious aerobic cocci and listeria monocytogenes.
Status
Terminated
Initiated
2020-12-02
Posted
not on file
Terminated
2023-01-03
Reason
Due to a customer report and internal investigation, it was confirmed that a portion of the affected product lot failed quality control (QC) for ATCC E. Faecalis 29212 versus daptomycin and it is a possibility to obtain false elevated MIC for clinical isolates with daptomycin even when quality control results are within specification.
Root cause
Process control
Action
During the week of 12/01/2020, the Recalling Firm is sending an "URGENT MEDICAL DEVICE RECALL" Letter via e-mail and/or mail to customers informing them that due to a customer report and internal investigation, it was confirmed that a portion of the affected product lot failed quality control (QC) for ATCC E. Faecalis 29212 versus daptomycin and it is a possibility to obtain false elevated MIC for clinical isolates with daptomycin even when quality control results are within specification. Customer are instructed to: 1. Discontinue use and discard, per their laboratory procedure/guidelines, any remaining inventory of the affected lot. 2. The laboratory should consider reviewing and confirming organism identification and antimicrobial susceptibility test results of stains yielding results suggestive of a daptomycin "non-susceptible" category. 3. They should retain their inventory of other B1017-216 panel lots as they are not impacted by this issue. 4. Share this information with their laboratory staff and retain the notification as part of their laboratory Quality System documentation. If they have forwarded any of the affected product to any additional laboratories, they are instructed to provide the information/notification. 5. To assure receipt of the notification/important communication, customers are instructed to respond within 10 days in one of the following ways: -Electronically, if they received the communication via email -Manually complete and return the enclosed Response Form. The Recalling Firm is investigating the root cause of the issues and appropriate actions to prevent recurrence will be implemented. For questions, contact Customer Support Center at: -websited http://www.beckmancoulter.com -Phone 1-800-677-7226 in the United States -For customers in the United States, if replacement product is needed: --Complete the attached "Replacement Order Form" and email to askbeckman@ Beckman.com or fax to 866-294-7850 --Call Client Servic
Quantity
685 panel boxes (20 panels per box) = 13,700 panels
Distribution
U.S. Nationwide distribution including in the states of AL, AZ, CA, FL, IN, KY, MA, MI, MO, NC, NH, NY, OH, OK, PA, SC, SD, TN, TX, VA, and WI. The countries of Australia, Austria, Bahamas, Bangladesh, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, Colombia, Croatia, Cura¿ao,¿St¿Eus, Czech¿Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, India, Iran, Ireland, Israel, Italy, Japan, Kenya, Kuwait, Latvia, Libya, Lithuania, Malaysia, Mexico, Netherlands, New Zealand, Norway, Oman, P.R. China, Pakistan, Philippines, Poland, Portugal, Qatar, Republic Korea, Romania, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, U.A.E., United Kingdom, Uruguay, Vatican City, Vietnam.
Lot, serial or model codes
Catalog Number: B1017-216 UDI Code: (01)15099590657550(17)210610(11)200610(10)2021-06-10  Lot Number: 2021-06-10
510(k) numbers
["K043545"]
PMA numbers
not on file
Product code
LRG
Firm FEI
2050012
Firm city
Brea
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register