Skip to the content

FDA Medical Device Recalls (openFDA)

Rubicor Medical: Rubicor EnCapsule Breast Biopsy Device, Part Number 31358, Product is manufactured and distributed by Rubicor Medical, Inc., 600 Chesapeake Dr., Redwood City, CA 94063

Recall number
Z-0715-06
Recalling firm
Rubicor Medical
Product
Rubicor EnCapsule Breast Biopsy Device, Part Number 31358, Product is manufactured and distributed by Rubicor Medical, Inc., 600 Chesapeake Dr., Redwood City, CA 94063
Status
Terminated
Initiated
2006-02-17
Posted
2006-04-11
Terminated
2006-05-11
Reason
Devices' sterility may be compromised due to lack of adequate sterilization.
Root cause
Other
Action
On 2/17/06, the firm initiated the recall and its notification was via sales rep retrieval and follow-up letter dated 3/13/06 explaining the reason for the recall and requesting the product be returned.
Quantity
19 units
Distribution
Nationwide. The product was released for distribution to 19 consignees in US. The firm''s sales force were able to retrieve 18 units, only one (1) was used by a consignee. The recall was appropriately extended to the consumer/user level; i.e., the wholesalers/distributors, pharmacists, physicians who received the recalled product. There were no shipments to U.S. Government facilities or foreign or Canadian distribution.
Lot, serial or model codes
Lot number: 06011817
510(k) numbers
["K020047","K023601","K030049","K032584","K052506"]
PMA numbers
not on file
Product code
GEI
Firm FEI
3004014191
Firm city
Redwood City
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register