FDA Medical Device Recalls (openFDA)

Abbott Laboratories: Abbott Laboratories, ARCHITECT 25-OH Vitamin D Reagent Kit, 03L52-25 (ROW) and 03L52-27 (U.S.) (1 x 100 tests), 03L52-35 (ROW) and 03L52-37 (U.S.) (5 x 100 tests); in vitro diagnostic.

Recall number
Z-0698-2013
Recalling firm
Abbott Laboratories
Product
Abbott Laboratories, ARCHITECT 25-OH Vitamin D Reagent Kit, 03L52-25 (ROW) and 03L52-27 (U.S.) (1 x 100 tests), 03L52-35 (ROW) and 03L52-37 (U.S.) (5 x 100 tests); in vitro diagnostic.
Status
Terminated
Initiated
2012-12-12
Posted
2013-01-23
Terminated
2016-12-23
Reason
The Architect 25-OH Vitamin D assay has the potential to generate falsely elevated or falsely depressed results when used with certain lots of Architect Reaction Vessels.
Root cause
Mixed-up of materials/components
Action
Abbott Laboratories sent a "Product Correction Immediate Action Required" letter dated December 12, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The customers were instructed to immediately discontinue use and destroy any remaining inventory of the reaction vessel lots listed; to clear the ARCHITECT RV hopper of impacted RV lots following the instructions provided, prior to continued use of the ARCHITECT 25-OH Vitamin D assay; to review the patient results obtained with the affected lots with the treating physician or heath care provider to ensure they are aware of the potential analytical issue and can determine whether re-evaluation of the patient is necessary; to forward a copy of the letter to any laboratory to whom they may have forwarded the reagent; and to retain a copy of the letter for their records and to complete and return the Customer Reply Form via fax to: 1-800-777-0051 For further questions please call your local Customer Service representative at 877-422-2688.
Quantity
28,600 total
Distribution
Worldwide Distribution -- USA (nationwide) and the countries of Austria, Barbados, Belgium, Canada, China, Colombia, Cyprus, Czech Republic, Denmark, France, Germany, Greece, Hungary, India, Ireland, Italy, Japan, Latvia, Libya, Malaysia, Netherlands, New Zealand, Poland, Portugal, Russia, Saudi Arabia, Serbia, Slovakia, Spain, Sweden, Switzerland, Turkey and United Kingdom.
Lot, serial or model codes
ARCHITECT 25-OH Vitamin D Reagent Kit list 3L52, lot numbers 02412C000, exp. 12/26/2012; 02512C000, exp. 12/26/2012; 02012G000, exp. 7/16/2013; 01612G000, exp. 7/17/13; 00912G000, exp. 7/10/13; 02512I000, exp. 10/8/2013; 02612I000, exp. 10/8/2013; 03512I000, exp. 10/10/2013; 03712I000, exp. 10/10/13; when used with ARCHITECT Reaction Vessel (RV) list 7C15-01, box lot numbers (bag lot numbers): 19320P100 (6F9891501); 19344P100 (4N5891091); 19404P100 (1J1948151);  19322P100 (LM3891731); 19373P100 (6F9891521); 19407P100 (4N5948361); 19324P100 (NK8891921); 19374P100 (JG7891631); 19408P100 (5P4948471); 19340P100 (1J1890911); 19375P100 (LM3891741); 19497P100 (2L2948281); 19341P100 (2L2891001); 19379P100 (NK8891931); 20026P100 (LM3947841)
510(k) numbers
K110619
PMA numbers
not on file
Product code
MRG
Firm FEI
1415939
Firm city
Abbott Park
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register