FDA Medical Device Recalls (openFDA)
Perkinelmer: NEOGRAM AMINO ACIDS AND ACYLCARNITINES TANDEM MASS SPECTROMETRY KIT Product Number: 3026-0030. This kit is intended for the measurement and evaluation of amino acid, free carnitine, and acylcarnitine concentrations from newborn heel prick blood samples dried on filter paper. (Table 1 of Attachment 2) details the analytes measured by the kit. Quantitative analysis of amino acids, free carnitine and acylcarnitines and their relationship with each other is intended to provide analyte concentration profiles that may aid in the screening of newborns for one or more of several metabolic disorders. This kit is to be used for in vitro diagnostic use only, by trained, qualified laboratory personnel.
- Recall number
- Z-0686-2013
- Recalling firm
- Perkinelmer
- Product
- NEOGRAM AMINO ACIDS AND ACYLCARNITINES TANDEM MASS SPECTROMETRY KIT Product Number: 3026-0030. This kit is intended for the measurement and evaluation of amino acid, free carnitine, and acylcarnitine concentrations from newborn heel prick blood samples dried on filter paper. (Table 1 of Attachment 2) details the analytes measured by the kit. Quantitative analysis of amino acids, free carnitine and acylcarnitines and their relationship with each other is intended to provide analyte concentration profiles that may aid in the screening of newborns for one or more of several metabolic disorders. This kit is to be used for in vitro diagnostic use only, by trained, qualified laboratory personnel.
- Status
- Terminated
- Initiated
- 2012-11-26
- Posted
- 2013-01-16
- Terminated
- 2014-03-07
- Reason
- Truncated V-Bottomed Microplate Wells used in assays contain defective wells.
- Root cause
- Process control
- Action
- Perkin Elmer initiated the recall on November 23, 2012. The Recall Letter will be delivered to the end customer by email facilitated through our PerkinElmer Technical Support group located in Akron Ohio. Technical Support will be asked to make sure the customers receive the Recall Letter and implement the recommended precautions: - Inspect the wells of row A and row H of the affected microplate lots for possible holes and cracks prior to use. - If a hole or crack is visually detected in a well of the microplate, the well should not be used for testing. - If a hole or crack is not visually detected in a well of the plate, it is required during use of the well, the volume of solution in the well should be visually confirmed to be consistent before and after incubation. Please contact PerkinEimer Technical Support for further information (800) 321 -9632, select prompt 2.
- Quantity
- 180 kits (20 plates/kit)
- Distribution
- Nationwide Distribution.
- Lot, serial or model codes
- Lot Numbers: 617341, 619236
- 510(k) numbers
- ["K021541"]
- PMA numbers
- not on file
- Product code
- JNB
- Firm FEI
- 3006159931
- Firm city
- Waltham
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28