Skip to the content

FDA Medical Device Recalls (openFDA)

Abbott Laboratories HPD/ADD: CA 15-3 Calibrators, list 9C08-01; the pack contains six 4-mL bottles of 115D8:DF3 reactive determinants prediluted in TRIS buffer with protein stabilizers to yield the following assay values: A - 0 u/mL, B - 15 U/mL, C - 60 U/mL, D - 120 U/mL, E - 180 U/mL and F - 250 U/mL; Abbott Laboratories, Abbott Park, IL 60064

Recall number
Z-0680-03
Recalling firm
Abbott Laboratories HPD/ADD
Product
CA 15-3 Calibrators, list 9C08-01; the pack contains six 4-mL bottles of 115D8:DF3 reactive determinants prediluted in TRIS buffer with protein stabilizers to yield the following assay values: A - 0 u/mL, B - 15 U/mL, C - 60 U/mL, D - 120 U/mL, E - 180 U/mL and F - 250 U/mL; Abbott Laboratories, Abbott Park, IL 60064
Status
Terminated
Initiated
2003-03-04
Posted
2003-03-26
Terminated
2004-09-15
Reason
The CA 15-3 Calibrators may cause depressed values for the CA 15-3 High Control and for patient specimens.
Root cause
Other
Action
Recalled by letter dated 3/4/03. The accounts were informed of the depressed values for the CA 15-3 High Control and for patient specimens, and were requested to discontinue use and destroy any of the three affected lots found in inventory.
Quantity
792 packs
Distribution
Nationwide and Ecuador, England, Hong Kong, Chile, Japan, Canada, Singapore, Australia, Pakistan, Argentina, Thailand, Aruba, Korea and Jamaica.
Lot, serial or model codes
list 9C08-01, lot 88127M100
510(k) numbers
["K963926","K964407"]
PMA numbers
not on file
Product code
MOI
Firm FEI
1415939
Firm city
Abbott Park
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register