Skip to the content

FDA Medical Device Recalls (openFDA)

ICU Medical, Inc.: ChemoLock Close Vial Spike REF CL-80 The Chemolock Closed System Transfer Device prevents the transfer of environmental contaminants. The Chemolock is needle free and cannot be deactivated, which will passively aid in preventing needlestick injuries and the exposure to cytotoxic medications for healthcare personnel.

Recall number
Z-0652-2025
Recalling firm
ICU Medical, Inc.
Product
ChemoLock Close Vial Spike REF CL-80 The Chemolock Closed System Transfer Device prevents the transfer of environmental contaminants. The Chemolock is needle free and cannot be deactivated, which will passively aid in preventing needlestick injuries and the exposure to cytotoxic medications for healthcare personnel.
Status
Open, Classified
Initiated
2024-11-06
Posted
2024-12-09
Terminated
not on file
Reason
Due to finished goods incorrectly packaged and labeled.
Root cause
Packaging process control
Action
On 11/06/2024, the firm send an "URGENT Medical Device Correction" Letter via third party (Sedgwick, Inc.) to customers informing them that product was incorrectly packaged and labeled. Product CL-80 package incorrectly contained product code CL-80S. The CL-80 is a used with a 20 mm or a 28 mm vial where as CL-80S is only use for a 20 mm vial. The issue is using the incorrect adapter may result in leakages. Actions to be taken by the Customer: Discontinue use mislabeled lot 13953938 of CL-80 and discard remaining product from this specific lot. 1. Check all inventory locations within your institution for the affected product listed in the notification and discontinue use. Quarantine and destroy all affected products following your institution s process for destruction. If destroying is not immediately possible at your facility, then the product should be quarantined until disposal. 2. Share this notification with all potential users of the device to ensure they are aware of this notification. If the devices are used at another location, please ensure this communication is delivered there./// 3. Complete and return the attached Customer Response Form to XXXX@sedgwick.com within 10 days of receipt to acknowledge your understanding of this notification./ / 4. DISTRIBUTORS: If you have distributed affected products to your customers, please immediately forward this notice to them. Request that they complete the response form and return it to XXXX@sedgwick.com. For questions/assistance, contact: Global Complaint Management 1-844-654-7780 (M-F, 8:00am 5:00pm CT) or ProductComplaintsPP@icumed.com To report adverse events or product complaints Customer Service 1-866-829-9025, option 8 or customerservice@icumed.com (M-F, 8:00 am 6:00 pm CT) Additional information or to request a credit. Field Action Processing 1-888-843-0245 xxx@sedgwick.com Questions about this communication
Quantity
6000 units
Distribution
Worldwide - U.S. Nationwide distribution in the state of IL and the country of Canada.
Lot, serial or model codes
Model/Catalog Number: CL-80 UDI code: (01)10887709057125(17)290401(30)50(10)13959938 Lot Number: 13959938
510(k) numbers
["K131549"]
PMA numbers
not on file
Product code
ONB
Firm FEI
2025816
Firm city
San Clemente
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register