FDA Medical Device Recalls (openFDA)
Genico, Inc., dba Genicon: Genicon 5mm Pyramidal Trocar, Sterile, Genion 6869 Stapoint Court, Suite 114 Winter Park, FL U.S.A. 32792, Rx Only. Catalog Numbers: 100-005-001 & 100-005-002. Used for incisions made for positioning the laparoscopic cannulas or ports.
- Recall number
- Z-0622-2009
- Recalling firm
- Genico, Inc., dba Genicon
- Product
- Genicon 5mm Pyramidal Trocar, Sterile, Genion 6869 Stapoint Court, Suite 114 Winter Park, FL U.S.A. 32792, Rx Only. Catalog Numbers: 100-005-001 & 100-005-002. Used for incisions made for positioning the laparoscopic cannulas or ports.
- Status
- Terminated
- Initiated
- 2008-09-25
- Posted
- 2008-12-24
- Terminated
- 2009-01-09
- Reason
- The stainless steel tips on the Genicon 5mm Pyramidal Trocar exhibited contamination that resembled rust and there was also pitting on the tip of the trocar.
- Root cause
- Manufacturing material removal
- Action
- Consignees were notified with a letter, dated 9/19/2008, by e-mail and fax on 09/25/2008. The letter stated a description of the issue, analysis and reliability data, indications of Device Malfunction, and recommendations. There was also a request for all defective devices to be returned to the distributor or Genicon for replacement. Genicon Customer service was to be contacted at 407-657-4851.
- Quantity
- 2,965 units (1340 OUS and 1625 US)
- Distribution
- Worldwide Distribution --- including USA and countries of Africa, United Kingdom, Denmark, Colombia, Belgium, Italy, Finland, Hong Kong, and Switzerland.
- Lot, serial or model codes
- Lot # I1219-K, exp: 11/01/2012; I1220-K, exp: 11/01/2012; I1278-K, exp: 01/01/2013; I1281-K, exp: 01/01/2013; I1393-K, exp: 03/01/2013; I1394-K, exp: 03/01/2013; I1415-K, exp: 03/01/2013; I1423-K, exp: 03/01/2013; I1487-K, exp: 04/01/2013; I1470-K, exp: 04/01/2013; and I1471-K, exp: 04/01/2013.
- 510(k) numbers
- K982472
- PMA numbers
- not on file
- Product code
- GEA
- Firm FEI
- 3002590791
- Firm city
- Winter Park
- Firm state
- FL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28