Greatbatch Medical: Biosense Webster MobiCath Bi-Directional Guiding Sheath. Small Curve catalog no. D140010, Large Curve catalog no. D140011. The MobiCath Bi-Directional Guiding Sheath is a flexible tipped percutaneous catheter introducer (or guiding sheath) designed for gaining access to the coronary systems. The device features adjustable tip geometry through use of a rotating handle to deflect the catheter. This catheter is assembled with a hemostasis valve and a sideport infusion line. The MobiCath Bi-Directional Guiding Sheath is intended for introducing various cardiovascular catheters into the heart, including the left side of the heart through the interatrial septum.

Recall number
Z-0611-2013
Recalling firm
Greatbatch Medical
Product
Biosense Webster MobiCath Bi-Directional Guiding Sheath. Small Curve catalog no. D140010, Large Curve catalog no. D140011. The MobiCath Bi-Directional Guiding Sheath is a flexible tipped percutaneous catheter introducer (or guiding sheath) designed for gaining access to the coronary systems. The device features adjustable tip geometry through use of a rotating handle to deflect the catheter. This catheter is assembled with a hemostasis valve and a sideport infusion line. The MobiCath Bi-Directional Guiding Sheath is intended for introducing various cardiovascular catheters into the heart, including the left side of the heart through the interatrial septum.
Status
Terminated
Initiated
2012-11-12
Posted
2012-12-22
Terminated
2014-05-15
Reason
Greatbatch Medical observed some anomalies of the inner lumen of the MobiCath Bi-Directional Guiding Sheaths upon implementation of a new visual inspection tool. The anomalies may include loose, string-like liner material, liner abrasion (fraying), and/or attached string-like liner material (from scratches or damage on the inner wall of sheath lumen).
Root cause
Nonconforming Material/Component
Action
Biosense Webster sent an "Urgent Field Safety Notice" dated November 12, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. A Voluntary Field Removal Certification Form was attached for customers to complete and return. Contact the firm at 800-729-9010 for questions related to this issue.
Quantity
10,201 units
Distribution
Worldwide Distribution-USA including the states of AL, AK, AZ, ALR, CA, CO, CT, DC, FL, GA, HI, IL IN, IA, KY, MD, MA, MI, MN, MO, NE, NV,NM, NJ, NH, NY, NC, OH, OK OR, PA, RI,SC,SD, TN, TX, UT,VT, VA, WA, and WI, and the countries of AUSTRIA, BELGIUM, SWITZERLAND, GERMANY, DENMARK, SPAIN, FINLAND, FRANCE, UNITED KINGDOM, GREECE, HUNGARY, IRELAND, ISRAEL, ITALY, THE NETHERLAND, NORWAY, SWEDEN, AUSTRALIA, and CANADA
Lot, serial or model codes
LOT: R1701247 , R1713591 , W2046680 , W2046681 , W2072907 , W2078460 , W2084872 , W2098771 , W2102124 , W2102133 , W2102134 , W2108295 , W2108296 , W2108298 , W2108299 , W2125688 , W2125690 , W2132584 , W2132588 , W2132590 , W2138965 , W2143147 , W2143149 , W2143152 , W2145645 , W2145677 , W2145680 , W2145681 , W2151287 , W2151305 , W2151308 , W2156746 , W2156748 , W2156750 , W2156751 , W2162496 , W2162497 , W2162498 , W2162499 , W2167986 , W2167987 , W2167988 , W2167990 , W2172705 , W2172706 , W2172707 , W2172708 , W2172710 , W2172712 , W2177727 , W2177728 , W2177733 , W2177734 , W2184476 , W2184483 , W2184484 , W2184487 , W2190457 , W2190461 , W2190464 , W2190470 , W2196349 , W2196351 , W2196352 , W2196355 , W2196359 , W2196361 , W2196363 , W2211842 , W2211844 , W2211846 , W2211849 , W2211852 , W2211853 , W2213887 , W2229974 , W2236588.
510(k) numbers
["K101784"]
PMA numbers
not on file
Product code
DYB
Firm FEI
2183787
Firm city
Minneapolis
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register