FDA Medical Device Recalls (openFDA)
Philips North America, LLC: HeartStart XL+ Defibrillator/Monitor, Model 861290
- Recall number
- Z-0608-2020
- Recalling firm
- Philips North America, LLC
- Product
- HeartStart XL+ Defibrillator/Monitor, Model 861290
- Status
- Terminated
- Initiated
- 2019-10-03
- Posted
- not on file
- Terminated
- 2023-04-19
- Reason
- Device may fail to turn on or unexpectedly attempt to restart, rendering it unable to return to a ready for use state. Additionally, this issue may occur when the device is in standby mode, when attempting to power on to run a self-test. If this occurs, the device will indicate that it is not ready for use.
- Root cause
- Process control
- Action
- Philips Healthcare issued Medical Device Correction Notices dated 10/3/2019 to customers via certified mail. Customers are advised to take the following actions: - The device is safe to use and can remain in service if it does not exhibit any of these behaviors described in this notice. - If you identify a device that exhibit these behaviors, please remove the device from service and contact Philips to request device. - Fill out, sign and send the reply form that is provided as the last page of the letter.
- Quantity
- 24,738 units
- Distribution
- US Nationwide
- Lot, serial or model codes
- All Units, Model number 861290
- 510(k) numbers
- ["K110825"]
- PMA numbers
- not on file
- Product code
- MKJ
- Firm FEI
- 1218950
- Firm city
- Andover
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28