Skip to the content

FDA Medical Device Recalls (openFDA)

Philips North America, LLC: HeartStart XL+ Defibrillator/Monitor, Model 861290

Recall number
Z-0608-2020
Recalling firm
Philips North America, LLC
Product
HeartStart XL+ Defibrillator/Monitor, Model 861290
Status
Terminated
Initiated
2019-10-03
Posted
not on file
Terminated
2023-04-19
Reason
Device may fail to turn on or unexpectedly attempt to restart, rendering it unable to return to a ready for use state. Additionally, this issue may occur when the device is in standby mode, when attempting to power on to run a self-test. If this occurs, the device will indicate that it is not ready for use.
Root cause
Process control
Action
Philips Healthcare issued Medical Device Correction Notices dated 10/3/2019 to customers via certified mail. Customers are advised to take the following actions: - The device is safe to use and can remain in service if it does not exhibit any of these behaviors described in this notice. - If you identify a device that exhibit these behaviors, please remove the device from service and contact Philips to request device. - Fill out, sign and send the reply form that is provided as the last page of the letter.
Quantity
24,738 units
Distribution
US Nationwide
Lot, serial or model codes
All Units, Model number 861290
510(k) numbers
["K110825"]
PMA numbers
not on file
Product code
MKJ
Firm FEI
1218950
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register