Skip to the content

FDA Medical Device Recalls (openFDA)

Philips North America: Azurion R3.0. Model Number 722234. Commercial Name: Azurion 7 M20. INTERVENTIONAL FLUOROSCOPIC XRAY SYSTEM

Recall number
Z-0597-2026
Recalling firm
Philips North America
Product
Azurion R3.0. Model Number 722234. Commercial Name: Azurion 7 M20. INTERVENTIONAL FLUOROSCOPIC XRAY SYSTEM
Status
Open, Classified
Initiated
2025-10-27
Posted
2025-12-09
Terminated
not on file
Reason
Between July 2024 and June 2025, Philips Azurion R3.0 systems with a FlexArm stand were shipped without the Source-to-Skin Distance (SSD) Spacer. The SSD spacer is necessary to meet the FDA minimum source-skin distance requirement of 38 cm for standard fluoroscopy applications.
Root cause
Radiation Control for Health and Safety Act
Action
Philips will send a customer notification to the users of the affected systems to alert them of the issue and provide the actions that should be taken to prevent risks to patients. Philips will contact customers to schedule a time for a Philips Field Service Engineer (FSE) to visit their site and install the SSD spacer to resolve the issue.
Quantity
110
Distribution
U.S.
Lot, serial or model codes
Model Number 722234 Commercial Name: Azurion 7 M20 How to identify: Philips Reference C&R 2025-IGT-BST-016
510(k) numbers
K200917
PMA numbers
not on file
Product code
JAA
Firm FEI
3006648320
Firm city
Cambridge
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register