Skip to the content

FDA Medical Device Recalls (openFDA)

Cepheid: Xpert CT/NG Catalog GXCT/NG-120 Batch 1000049568, Catalog GXCT/NG-10 Batch 1000053075, Catalog GXCT/NG-10 Batch 1000056156 and Catalog GXCT/NG-10 Batch 1000056157. Is a qualitative in vitro real-time PCR test.

Recall number
Z-0592-2018
Recalling firm
Cepheid
Product
Xpert CT/NG Catalog GXCT/NG-120 Batch 1000049568, Catalog GXCT/NG-10 Batch 1000053075, Catalog GXCT/NG-10 Batch 1000056156 and Catalog GXCT/NG-10 Batch 1000056157. Is a qualitative in vitro real-time PCR test.
Status
Terminated
Initiated
2017-10-09
Posted
not on file
Terminated
2018-02-13
Reason
Cepheid has received reports that some of our customers have experienced a higher number of temperature drift errors than would typically be expected when using this test and its intended use specimen types. Investigation of the issue also revealed an occasional probability of false negative results which remains consistent with package insert claims.
Root cause
Equipment maintenance
Action
An Urgent Medical Device Recall letter (response form included), dated October 9, 2017, was sent to affected consignees. The letter identified the affected product, the issue involved, impact, and actions to be taken. Cepheid will provide replacement product for Xpert CT/NG assay product identified in this recall. Customers are to complete and the return the Response Form provided in order to receive replacement product. Additionally, the letter instructs users to review their laboratory statistics to see if their positivity rates are trending low during use of the affected lots. If so, they are to review the negative patient results according to their laboratory's Quality Management System. Customers are to refer to the applicable contact information provided if there are any questions.
Quantity
3,586 kits (49,400 units)
Distribution
Worldwide Distribution -- USA, Puerto Rico, and Haiti.
Lot, serial or model codes
Lot numbers: 25718, 25719, 25725, and 25727.
510(k) numbers
K121710
PMA numbers
not on file
Product code
LSL
Firm FEI
3004530258
Firm city
Sunnyvale
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register