FDA Medical Device Recalls (openFDA)
GE Healthcare, LLC: GE LOGIQ E10 Ultrasound System, Model #530000E10. The device is intended for use by a qualified physician for ultrasound evaluation of Fetal/Obstetrics; Abdominal; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculoskeletal Conventional and Superficial; Urology (including prostate); Transrectal; Transvaginal; Transesophageal and Intraoperative (vascular).
- Recall number
- Z-0588-2019
- Recalling firm
- GE Healthcare, LLC
- Product
- GE LOGIQ E10 Ultrasound System, Model #530000E10. The device is intended for use by a qualified physician for ultrasound evaluation of Fetal/Obstetrics; Abdominal; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculoskeletal Conventional and Superficial; Urology (including prostate); Transrectal; Transvaginal; Transesophageal and Intraoperative (vascular).
- Status
- Terminated
- Initiated
- 2018-09-14
- Posted
- not on file
- Terminated
- 2019-05-29
- Reason
- There is a potential for an undetected over temperature condition on the trans-esophageal (TEE) probe leading to a possible burn injury.
- Root cause
- Process change control
- Action
- The recalling firm, GE Healthcare, issued an "URGENT MEDICAL DEVICE CORRECTION" letter dated 9/14/2018 via traceable means to each of their customer locations. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: Do not use a TEE probe (GE model 6Tc-RS) from another device on the LOGIQ E10 system until your LOGIQ E10 is corrected. You may scan normally with any other probes. GE Healthcare will correct all affected products at no cost to you. A GE Healthcare representative will contact you to arrange for the correction. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
- Quantity
- 103 devices(64 US; 39 OUS)
- Distribution
- Worldwide Distribution: US (nationwide) to states of: AZ, IL, MA, ME, MN, MT, PA, TX, and WA and countries of: Foreign distribution was made to Austria, Denmark, Finland, and Norway.
- Lot, serial or model codes
- Serial numbers 500055US7, 500103US7, 500022US7, 500059US7, 00066US7, 500083US7, 500057US7, 500130US7, 500114US7, 500105US7, 500111US7, 500039US7,500132US7, 500064US7, 500128US7, 500131US7, 500063US7, 500062US7, 500104US7, 500139US7, 500126US7, 500127US7, 500125US7, 500138US7, 500140US7, 500095US7, 500099US7, 500100US7, 500102US7, 500097US7, 500060US7, 500020US7, 500002US7, 500019US7, 500093US7, 500003US7, 500004US7,500005US7, 500065US7, 500072US7, 500058US7, 500049US7, 500050US7, 500047US7, 500035US7, 500021US7, 500023US7, 500024US7, 500025US7, 500026US7, 500007US7, 500008US7, 500009US7, 500010US7, 500013US7, 500014US7, 500016US7, 500017US7, 500012US7, 500029US7, 500043US7, 500061US7, 500006US7, 500015US7, 500085US7, 500080US7, 500042US7, 500070US7, 500030US7, 500122US7, 500082US7, 500135US7, 500121US7, 500086US7, 500027US7, 500048US7, 500053US7, 500054US7, 500056US7, 500116US7, 500038US7, 500045US7, 500036US7, 500051US7, 500073US7, 500076US7, 500079US7, 500084US7, 500091US7, 500034US7, 500077US7, 500133US7, 500087US7, 500134US7, 500074US7, 500068US7, 500037US7, 500018US7, 500069US7, 500067US7, 500071US7, 500044US7, and 500090US7.
- 510(k) numbers
- K173555
- PMA numbers
- not on file
- Product code
- IYN
- Firm FEI
- 2126677
- Firm city
- Waukesha
- Firm state
- WI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- 3e6df46828c1
- Size
- 288213574 bytes
- Published
- 2026-10-06
- Read
- 2026-10-07T03:20:25.099Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-14
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 3e6df46828c1, published 2026-10-06