Skip to the content

FDA Medical Device Recalls (openFDA)

GE Healthcare, LLC: GE LOGIQ E10 Ultrasound System, Model #530000E10. The device is intended for use by a qualified physician for ultrasound evaluation of Fetal/Obstetrics; Abdominal; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculoskeletal Conventional and Superficial; Urology (including prostate); Transrectal; Transvaginal; Transesophageal and Intraoperative (vascular).

Recall number
Z-0588-2019
Recalling firm
GE Healthcare, LLC
Product
GE LOGIQ E10 Ultrasound System, Model #530000E10. The device is intended for use by a qualified physician for ultrasound evaluation of Fetal/Obstetrics; Abdominal; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculoskeletal Conventional and Superficial; Urology (including prostate); Transrectal; Transvaginal; Transesophageal and Intraoperative (vascular).
Status
Terminated
Initiated
2018-09-14
Posted
not on file
Terminated
2019-05-29
Reason
There is a potential for an undetected over temperature condition on the trans-esophageal (TEE) probe leading to a possible burn injury.
Root cause
Process change control
Action
The recalling firm, GE Healthcare, issued an "URGENT MEDICAL DEVICE CORRECTION" letter dated 9/14/2018 via traceable means to each of their customer locations. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: Do not use a TEE probe (GE model 6Tc-RS) from another device on the LOGIQ E10 system until your LOGIQ E10 is corrected. You may scan normally with any other probes. GE Healthcare will correct all affected products at no cost to you. A GE Healthcare representative will contact you to arrange for the correction. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Quantity
103 devices(64 US; 39 OUS)
Distribution
Worldwide Distribution: US (nationwide) to states of: AZ, IL, MA, ME, MN, MT, PA, TX, and WA and countries of: Foreign distribution was made to Austria, Denmark, Finland, and Norway.
Lot, serial or model codes
Serial numbers 500055US7, 500103US7, 500022US7, 500059US7, 00066US7, 500083US7, 500057US7, 500130US7, 500114US7, 500105US7, 500111US7, 500039US7,500132US7, 500064US7, 500128US7, 500131US7, 500063US7, 500062US7, 500104US7, 500139US7, 500126US7, 500127US7, 500125US7, 500138US7, 500140US7, 500095US7, 500099US7, 500100US7, 500102US7, 500097US7, 500060US7, 500020US7, 500002US7, 500019US7, 500093US7, 500003US7, 500004US7,500005US7, 500065US7, 500072US7, 500058US7, 500049US7, 500050US7, 500047US7, 500035US7, 500021US7, 500023US7, 500024US7, 500025US7, 500026US7, 500007US7, 500008US7, 500009US7, 500010US7, 500013US7, 500014US7, 500016US7, 500017US7, 500012US7, 500029US7, 500043US7, 500061US7, 500006US7, 500015US7, 500085US7, 500080US7, 500042US7, 500070US7, 500030US7, 500122US7, 500082US7, 500135US7, 500121US7, 500086US7, 500027US7, 500048US7, 500053US7, 500054US7, 500056US7, 500116US7, 500038US7, 500045US7, 500036US7, 500051US7, 500073US7, 500076US7, 500079US7, 500084US7, 500091US7, 500034US7, 500077US7, 500133US7, 500087US7, 500134US7, 500074US7, 500068US7, 500037US7, 500018US7, 500069US7, 500067US7, 500071US7, 500044US7, and 500090US7.
510(k) numbers
K173555
PMA numbers
not on file
Product code
IYN
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
3e6df46828c1
Size
288213574 bytes
Published
2026-10-06
Read
2026-10-07T03:20:25.099Z
Source file
Source file retained · verify
Refresh due
2026-10-14

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 3e6df46828c1, published 2026-10-06

Search this source