Fresenius Kabi USA, LLC: IVENIX INFUSION SYSTEM, LVP Primary Administration Set, Dual-Inlet, Low-Sorbing, Needle-Free Port, Dual Y-Site (Qty 25). Large Volume Infusion Pump Administration Set.
- Recall number
- Z-0586-2026
- Recalling firm
- Fresenius Kabi USA, LLC
- Product
- IVENIX INFUSION SYSTEM, LVP Primary Administration Set, Dual-Inlet, Low-Sorbing, Needle-Free Port, Dual Y-Site (Qty 25). Large Volume Infusion Pump Administration Set.
- Status
- Open, Classified
- Initiated
- 2025-11-03
- Posted
- 2025-12-03
- Terminated
- not on file
- Reason
- Incorrect assembly of specific lot of LVP Primary Administration Set that could result in serious patient injury.
- Root cause
- Process control
- Action
- An URGENT - Voluntary Recall notice dated 10/30/25 was emailed to consignees. Consignees are instructed to discontinue use and distribution of affected devices immediately; check inventory and quarantine all affected devices. The recall notification is to be shared with users and inventory replaced with unaffected stock. Consignees with no affected devices are to complete and return the provided response form. Consignees with affected devices are to either destroy or return affected devices; replacements can be requested by contacting Fresenius Kabi Customer Service at 855-354-6387 (x1). If consignees destroy product, Fresenius Kabi will provide a Certificate of Destruction which must be completed and returned. Consignees with any questions are to email Ivenix_support@fresenius-kabi.com or call 855-354-6387.
- Quantity
- 483 cases (12,075 eaches)
- Distribution
- US Nationwide distribution in the states of IL, MD, MI, OK, SC, TX, VA & WI.
- Lot, serial or model codes
- Product Code: SET-0032-25; UDI: 00811505030214; Lot No. FA25B03126.
- 510(k) numbers
- ["K183311"]
- PMA numbers
- not on file
- Product code
- FRN
- Firm FEI
- 3014732157
- Firm city
- North Andover
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28