Skip to the content

FDA Medical Device Recalls (openFDA)

Fresenius Kabi USA, LLC: IVENIX INFUSION SYSTEM, LVP Primary Administration Set, Dual-Inlet, Low-Sorbing, Needle-Free Port, Dual Y-Site (Qty 25). Large Volume Infusion Pump Administration Set.

Recall number
Z-0586-2026
Recalling firm
Fresenius Kabi USA, LLC
Product
IVENIX INFUSION SYSTEM, LVP Primary Administration Set, Dual-Inlet, Low-Sorbing, Needle-Free Port, Dual Y-Site (Qty 25). Large Volume Infusion Pump Administration Set.
Status
Open, Classified
Initiated
2025-11-03
Posted
2025-12-03
Terminated
not on file
Reason
Incorrect assembly of specific lot of LVP Primary Administration Set that could result in serious patient injury.
Root cause
Process control
Action
An URGENT - Voluntary Recall notice dated 10/30/25 was emailed to consignees. Consignees are instructed to discontinue use and distribution of affected devices immediately; check inventory and quarantine all affected devices. The recall notification is to be shared with users and inventory replaced with unaffected stock. Consignees with no affected devices are to complete and return the provided response form. Consignees with affected devices are to either destroy or return affected devices; replacements can be requested by contacting Fresenius Kabi Customer Service at 855-354-6387 (x1). If consignees destroy product, Fresenius Kabi will provide a Certificate of Destruction which must be completed and returned. Consignees with any questions are to email Ivenix_support@fresenius-kabi.com or call 855-354-6387.
Quantity
483 cases (12,075 eaches)
Distribution
US Nationwide distribution in the states of IL, MD, MI, OK, SC, TX, VA & WI.
Lot, serial or model codes
Product Code: SET-0032-25; UDI: 00811505030214; Lot No. FA25B03126.
510(k) numbers
["K183311"]
PMA numbers
not on file
Product code
FRN
Firm FEI
3014732157
Firm city
North Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register