Skip to the content

FDA Medical Device Recalls (openFDA)

Sybaritic, Inc: Dermosonic Model 430300 and 440100 Non-Invasive Subdermal Therapy System

Recall number
Z-0578-05
Recalling firm
Sybaritic, Inc
Product
Dermosonic Model 430300 and 440100 Non-Invasive Subdermal Therapy System
Status
Terminated
Initiated
2005-01-05
Posted
2005-03-09
Terminated
2007-02-07
Reason
Device is performing functions that are not cleared in the 510(k).
Root cause
Other
Action
The firm initiated the recall procedure 01/05/05. A phone call is made to each facility which has a dermosonic machine. The phone call serves the purpose of telling the customer that their dermosonic machine needs a programming chip replaced, the manual updated and a 'caution' sticker be placed on the machine. An employee of Computer Repair.com then contacts the customer to set up a time to replace the programming chip and manual, and place a 'caution' sticker on the machine.
Quantity
185 units
Distribution
AZ, CA, CO, CT, FL, GA, ID, IL, IA, KS, LA, MA, MI, MN, NV, NJ, NM, NY, NC, OH, OK, OR, SC, SD, TN, TX, UT, VA & WI and to the Phillipines
Lot, serial or model codes
serial numbers: 8, 960, 1341, 1343, 1344, 1351 through 1354, 1359, 1363, 1365 through 1368, 1372,  1379, 1581, 1583, 1584, 1589 through 1593, 1598, 1599, 1756, 1757, 1760, 1771,  1784, 1786, 1792, 1793, 2001, 2005, 2007, 2012 through 2018, 2020 through 2025, 2028, 2029, 2031, 2032, 2036, 2037, 2038, 2041, 2043, 2044, 2047,  2175, 2176, 2177, 2180 through 2187, 2189 through 2195, 2197, 2198, 2348, 2355 through 2370, 2372, 2373, 2374, 2375, 2376, 2378, 2381, 2382,  2389, 2390, 2392, 2393, 2400 through 2433, 2435, 2436, 2442, 2443, 2453, 2475, 2560,  2562, 2563, 2564, 2565, 2566, 2568, 2569, 2572 through 2579, 2590, 2592, 2593, 2595, 2596, 2597, 2598, and 2599
510(k) numbers
["K024307"]
PMA numbers
not on file
Product code
IMI
Firm FEI
2131125
Firm city
Bloomington
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register