FDA Medical Device Recalls (openFDA)
Olympus Corporation of the Americas: Brand Name: EZDilate Nephrostomy Balloon Catheter and Inflation Device Kit Product Name: Olympus Nephro-EZDilate Nephrostomy Balloon Catheter Model/Catalog Number: BPCN0815K Product Description: The Nephro-EZDilate nephrostomy balloon catheter is a reinforced catheter attached to a distal dilatation balloon. It has a radiopaque tip and two radiopaque markers positioned on the catheter inside the balloon that define the working length
- Recall number
- Z-0572-2026
- Recalling firm
- Olympus Corporation of the Americas
- Product
- Brand Name: EZDilate Nephrostomy Balloon Catheter and Inflation Device Kit Product Name: Olympus Nephro-EZDilate Nephrostomy Balloon Catheter Model/Catalog Number: BPCN0815K Product Description: The Nephro-EZDilate nephrostomy balloon catheter is a reinforced catheter attached to a distal dilatation balloon. It has a radiopaque tip and two radiopaque markers positioned on the catheter inside the balloon that define the working length
- Status
- Open, Classified
- Initiated
- 2025-10-17
- Posted
- 2025-11-20
- Terminated
- not on file
- Reason
- Device kits contained an incorrectly sized 4mm x 4cm urinary tract balloon catheter (BURS0404) instead of the correct 8mm x 15cm nephrostomy tract balloon catheter (BPCN0815). All other components in the kits were correctly included and packaged in the kit.
- Root cause
- Process control
- Action
- On October 17, 2025, "URGENT: MEDICAL DEVICE REMOVAL" letters were sent to customers. Actions required: 1. 1. Examine your inventory and quarantine any affected devices. 2. Cease usage of the product with immediate effect. 3. Return any remaining affected product, please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization. Olympus will issue a credit to your facility upon return of your affected product. 4. Please forward this notice to other users who may have the affected products if you have further distributed it. Olympus requests you to report any complaints to our Technical Assistance Center (TAC) at 1- 800-848-9024, option 1, and the FDA If you require additional information, please do not hesitate to contact directly by phone at (647) 999-3203 or by e-mail Cynthia.Ow@Olympus.com.
- Quantity
- 60 units
- Distribution
- US Domestic: AK, CA, IA, KY, WA, WI; OUS International: Canada and Germany.
- Lot, serial or model codes
- Model: BPCN0815K; UDI: 00821925033016; Lot Number: 25037006;
- 510(k) numbers
- K132383
- PMA numbers
- not on file
- Product code
- LJE
- Firm FEI
- 2429304
- Firm city
- Center Valley
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28