Skip to the content

FDA Medical Device Recalls (openFDA)

Olympus Corporation of the Americas: Brand Name: EZDilate Nephrostomy Balloon Catheter and Inflation Device Kit Product Name: Olympus Nephro-EZDilate Nephrostomy Balloon Catheter Model/Catalog Number: BPCN0815K Product Description: The Nephro-EZDilate nephrostomy balloon catheter is a reinforced catheter attached to a distal dilatation balloon. It has a radiopaque tip and two radiopaque markers positioned on the catheter inside the balloon that define the working length

Recall number
Z-0572-2026
Recalling firm
Olympus Corporation of the Americas
Product
Brand Name: EZDilate Nephrostomy Balloon Catheter and Inflation Device Kit Product Name: Olympus Nephro-EZDilate Nephrostomy Balloon Catheter Model/Catalog Number: BPCN0815K Product Description: The Nephro-EZDilate nephrostomy balloon catheter is a reinforced catheter attached to a distal dilatation balloon. It has a radiopaque tip and two radiopaque markers positioned on the catheter inside the balloon that define the working length
Status
Open, Classified
Initiated
2025-10-17
Posted
2025-11-20
Terminated
not on file
Reason
Device kits contained an incorrectly sized 4mm x 4cm urinary tract balloon catheter (BURS0404) instead of the correct 8mm x 15cm nephrostomy tract balloon catheter (BPCN0815). All other components in the kits were correctly included and packaged in the kit.
Root cause
Process control
Action
On October 17, 2025, "URGENT: MEDICAL DEVICE REMOVAL" letters were sent to customers. Actions required: 1. 1. Examine your inventory and quarantine any affected devices. 2. Cease usage of the product with immediate effect. 3. Return any remaining affected product, please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization. Olympus will issue a credit to your facility upon return of your affected product. 4. Please forward this notice to other users who may have the affected products if you have further distributed it. Olympus requests you to report any complaints to our Technical Assistance Center (TAC) at 1- 800-848-9024, option 1, and the FDA If you require additional information, please do not hesitate to contact directly by phone at (647) 999-3203 or by e-mail Cynthia.Ow@Olympus.com.
Quantity
60 units
Distribution
US Domestic: AK, CA, IA, KY, WA, WI; OUS International: Canada and Germany.
Lot, serial or model codes
Model: BPCN0815K; UDI: 00821925033016; Lot Number: 25037006;
510(k) numbers
K132383
PMA numbers
not on file
Product code
LJE
Firm FEI
2429304
Firm city
Center Valley
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register