Skip to the content

FDA Medical Device Recalls (openFDA)

Sterilmed Inc: Masks - Laryngeal Model #'s: AMB321-300-000 AMB321-400-000 AMB321-500-000 LMA12015 LMA12020 LMA12030 LMA12040 LMA12050 LMA12120 LMA12125 LMA12130 LMA12140 and LMA21140 An oropharyngeal airway is a device inserted into a patient's pharynx through the mouth to provide a patent airway.

Recall number
Z-0563-2009
Recalling firm
Sterilmed Inc
Product
Masks - Laryngeal Model #'s: AMB321-300-000 AMB321-400-000 AMB321-500-000 LMA12015 LMA12020 LMA12030 LMA12040 LMA12050 LMA12120 LMA12125 LMA12130 LMA12140 and LMA21140 An oropharyngeal airway is a device inserted into a patient's pharynx through the mouth to provide a patent airway.
Status
Terminated
Initiated
2008-08-08
Posted
2009-01-13
Terminated
2011-04-26
Reason
SteriMed, Inc is conducting a product removal of a small group of devices individually packaged and sealed within a pouch. The removal is being conducted because the possibility exists that some of the pouches were not properly sealed on one end. A breach in packaging seal or a failure in packaging integrity has the possibility of risk to the patient in terms of transmitting organisms capable
Root cause
Packaging process control
Action
Consignees were notified of the recall via verbal communication in person by on-site technician and or their local sales representative on 8/8/08. Others were contacted via fax or e-mail followed by on-site visits 8/11/08. SteriMed also provided consignees (Materials Manager/Risk Manager) with a "Urgent Product Removal" recall letter (August 14, 2008) in person informing them of the potential of receiving affected product. Included in the Product Removal Notification was a Customer Reply Form to inventory quantity of product received and quantity returned. Contact SterilMed, Inc. at 1-888-541-0078 for assistance.
Quantity
124
Distribution
AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, MA, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, WA, WI, and WV OUS: CANADA
Lot, serial or model codes
n/a
510(k) numbers
not on file
PMA numbers
not on file
Product code
CAE
Firm FEI
3001237663
Firm city
Osseo
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register