FDA Medical Device Recalls (openFDA)
Sterilmed Inc: Burrs Model #'s: ANSL-3SB ANSL-5D ANSQD11-4OVDC HAL5056-118 HAL5056-166 HAL5091-106 HAL5091-112 HAL5091-122 HAL5091-125 HAL5091-136 HAL5091-144 HAL5091-164 HAL5092-120 HAL5092-168 HAL5092-170 KOM04-D0139 LIN5091-101 LIN5091-102 LIN5091-142 LIN5092-132 LIN5092-160 MED874-447 MICZB-228 MICZB-232 MID14CY50 MID14MH30D MIDG6-450 MIDG8-130 MIDMC254 STR1607-2-35 STR1608-6-137 STR1608-6-141 STR1608-6-149 STR277-10-62 STR5120-10-40 STR5120-10-50 STR5120-71-48 STR5300-10-511 STR5300-10-513 STR5300-10-903 STR5300-20-513 STR5300-30-903 XOM31-55641 ZIM5056-116 ZIM5056-117 ZIM5056-118 ZIM5056-125 ZIM5056-140 ZIM5056-150 ZIM5056-159 ZIM5056-161 ZIM5056-162 ZIM5056-164 and ZIM5056-167 Surgical instruments motors and accessories are AC-powered, battery-powered, or air-powered devices intended for use during surgical procedures to provide power to operate various accessories or attachments to cut hard tissue or bone and soft tissue.
- Recall number
- Z-0550-2009
- Recalling firm
- Sterilmed Inc
- Product
- Burrs Model #'s: ANSL-3SB ANSL-5D ANSQD11-4OVDC HAL5056-118 HAL5056-166 HAL5091-106 HAL5091-112 HAL5091-122 HAL5091-125 HAL5091-136 HAL5091-144 HAL5091-164 HAL5092-120 HAL5092-168 HAL5092-170 KOM04-D0139 LIN5091-101 LIN5091-102 LIN5091-142 LIN5092-132 LIN5092-160 MED874-447 MICZB-228 MICZB-232 MID14CY50 MID14MH30D MIDG6-450 MIDG8-130 MIDMC254 STR1607-2-35 STR1608-6-137 STR1608-6-141 STR1608-6-149 STR277-10-62 STR5120-10-40 STR5120-10-50 STR5120-71-48 STR5300-10-511 STR5300-10-513 STR5300-10-903 STR5300-20-513 STR5300-30-903 XOM31-55641 ZIM5056-116 ZIM5056-117 ZIM5056-118 ZIM5056-125 ZIM5056-140 ZIM5056-150 ZIM5056-159 ZIM5056-161 ZIM5056-162 ZIM5056-164 and ZIM5056-167 Surgical instruments motors and accessories are AC-powered, battery-powered, or air-powered devices intended for use during surgical procedures to provide power to operate various accessories or attachments to cut hard tissue or bone and soft tissue.
- Status
- Terminated
- Initiated
- 2008-08-08
- Posted
- 2009-01-13
- Terminated
- 2011-04-26
- Reason
- SteriMed, Inc is conducting a product removal of a small group of devices individually packaged and sealed within a pouch. The removal is being conducted because the possibility exists that some of the pouches were not properly sealed on one end. A breach in packaging seal or a failure in packaging integrity has the possibility of risk to the patient in terms of transmitting organisms capable
- Root cause
- Packaging process control
- Action
- Consignees were notified of the recall via verbal communication in person by on-site technician and or their local sales representative on 8/8/08. Others were contacted via fax or e-mail followed by on-site visits 8/11/08. SteriMed also provided consignees (Materials Manager/Risk Manager) with a "Urgent Product Removal" recall letter (August 14, 2008) in person informing them of the potential of receiving affected product. Included in the Product Removal Notification was a Customer Reply Form to inventory quantity of product received and quantity returned. Contact SterilMed, Inc. at 1-888-541-0078 for assistance.
- Quantity
- 125
- Distribution
- AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, MA, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, WA, WI, and WV OUS: CANADA
- Lot, serial or model codes
- n/a
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- GFF
- Firm FEI
- 3001237663
- Firm city
- Osseo
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28