Skip to the content

FDA Medical Device Recalls (openFDA)

Sterilmed Inc: Burrs Model #'s: ANSL-3SB ANSL-5D ANSQD11-4OVDC HAL5056-118 HAL5056-166 HAL5091-106 HAL5091-112 HAL5091-122 HAL5091-125 HAL5091-136 HAL5091-144 HAL5091-164 HAL5092-120 HAL5092-168 HAL5092-170 KOM04-D0139 LIN5091-101 LIN5091-102 LIN5091-142 LIN5092-132 LIN5092-160 MED874-447 MICZB-228 MICZB-232 MID14CY50 MID14MH30D MIDG6-450 MIDG8-130 MIDMC254 STR1607-2-35 STR1608-6-137 STR1608-6-141 STR1608-6-149 STR277-10-62 STR5120-10-40 STR5120-10-50 STR5120-71-48 STR5300-10-511 STR5300-10-513 STR5300-10-903 STR5300-20-513 STR5300-30-903 XOM31-55641 ZIM5056-116 ZIM5056-117 ZIM5056-118 ZIM5056-125 ZIM5056-140 ZIM5056-150 ZIM5056-159 ZIM5056-161 ZIM5056-162 ZIM5056-164 and ZIM5056-167 Surgical instruments motors and accessories are AC-powered, battery-powered, or air-powered devices intended for use during surgical procedures to provide power to operate various accessories or attachments to cut hard tissue or bone and soft tissue.

Recall number
Z-0550-2009
Recalling firm
Sterilmed Inc
Product
Burrs Model #'s: ANSL-3SB ANSL-5D ANSQD11-4OVDC HAL5056-118 HAL5056-166 HAL5091-106 HAL5091-112 HAL5091-122 HAL5091-125 HAL5091-136 HAL5091-144 HAL5091-164 HAL5092-120 HAL5092-168 HAL5092-170 KOM04-D0139 LIN5091-101 LIN5091-102 LIN5091-142 LIN5092-132 LIN5092-160 MED874-447 MICZB-228 MICZB-232 MID14CY50 MID14MH30D MIDG6-450 MIDG8-130 MIDMC254 STR1607-2-35 STR1608-6-137 STR1608-6-141 STR1608-6-149 STR277-10-62 STR5120-10-40 STR5120-10-50 STR5120-71-48 STR5300-10-511 STR5300-10-513 STR5300-10-903 STR5300-20-513 STR5300-30-903 XOM31-55641 ZIM5056-116 ZIM5056-117 ZIM5056-118 ZIM5056-125 ZIM5056-140 ZIM5056-150 ZIM5056-159 ZIM5056-161 ZIM5056-162 ZIM5056-164 and ZIM5056-167 Surgical instruments motors and accessories are AC-powered, battery-powered, or air-powered devices intended for use during surgical procedures to provide power to operate various accessories or attachments to cut hard tissue or bone and soft tissue.
Status
Terminated
Initiated
2008-08-08
Posted
2009-01-13
Terminated
2011-04-26
Reason
SteriMed, Inc is conducting a product removal of a small group of devices individually packaged and sealed within a pouch. The removal is being conducted because the possibility exists that some of the pouches were not properly sealed on one end. A breach in packaging seal or a failure in packaging integrity has the possibility of risk to the patient in terms of transmitting organisms capable
Root cause
Packaging process control
Action
Consignees were notified of the recall via verbal communication in person by on-site technician and or their local sales representative on 8/8/08. Others were contacted via fax or e-mail followed by on-site visits 8/11/08. SteriMed also provided consignees (Materials Manager/Risk Manager) with a "Urgent Product Removal" recall letter (August 14, 2008) in person informing them of the potential of receiving affected product. Included in the Product Removal Notification was a Customer Reply Form to inventory quantity of product received and quantity returned. Contact SterilMed, Inc. at 1-888-541-0078 for assistance.
Quantity
125
Distribution
AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, MA, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, WA, WI, and WV OUS: CANADA
Lot, serial or model codes
n/a
510(k) numbers
not on file
PMA numbers
not on file
Product code
GFF
Firm FEI
3001237663
Firm city
Osseo
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register