Skip to the content

FDA Medical Device Recalls (openFDA)

Immuno-Mycologics, Inc: CRYPTOCOCCAL ANTIGEN LATERAL FLOW ASSAY, IMMY, Lateral Flow Assay for the Detection of Cryptococcal Antigen in serum and CSF, REF CR2003, IVD, UDI: (01)00816387020919

Recall number
Z-0545-2022
Recalling firm
Immuno-Mycologics, Inc
Product
CRYPTOCOCCAL ANTIGEN LATERAL FLOW ASSAY, IMMY, Lateral Flow Assay for the Detection of Cryptococcal Antigen in serum and CSF, REF CR2003, IVD, UDI: (01)00816387020919
Status
Terminated
Initiated
2022-01-03
Posted
not on file
Terminated
2024-05-15
Reason
The firm found immunoassay products used to detect cryptococcal antigen in serum and cerebrospinal fluid to have reduced specificity (90% now versus 99% before). As a result, a small number of samples with positive test results may be false positives, which may cause some patients to initiate unnecessary anti-fungal therapy. Samples with negative test results are NOT affected.
Root cause
Material/Component Contamination
Action
Customers who received product were contacted via letter Urgent: Medical Device Recall, Cryptococcal Antigen Lateral Flow Assay, dated January 3, 2022, beginning January 3, 2022, to locate and immediately destroy their affected devices. A recall notification and response form was attached to the email that is to be completed by the customers and returned to the firm.
Quantity
411 kits
Distribution
US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IL, IN, KY, LA, MA, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NY, OH, OR, PA, RI, SC, TN, TX, UT, VA, WA.
Lot, serial or model codes
F1021035 and F1021036
510(k) numbers
["K112422"]
PMA numbers
not on file
Product code
GMD
Firm FEI
1627497
Firm city
Norman
Firm state
OK
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register