ASI Medical, Inc.: SEAC Advanced Dental System, Model A591561 The SEAC Advanced Dental System is a self-contained dental delivery system with a built-in air compressor,vacuum system and water supply. It is designed operate dental handpiece attachments and suction instruments. The SEAC Advanced Dental System can also be upgraded to include a Micadent air abrasion system mounted inside the unit with the handpiece mounted in a holder along side the other handpiece connections. The Micadent is used to perform intra oral air abrasion procedures including etching and preparing surfaces for composite restoration. The SEAC Advanced Dental System is substantially similar in design and functions as the SEAC Mobile Dental System manufactured by ASI Medical, Inc. The Micadent is manufactured by Medidents and will be supplied to ASIas a completed product. The Micadent is substantially the same as it is provided and is attached to an accessory air line inside the SEAC unit for operation. It is intended for use to provide general dental restorative care and hygiene procedures in both traditional dental office settings and/or mobile applications by or under the supervision of a licensed dentist or a hygienist if permitted by applicable law.

Recall number
Z-0545-2014
Recalling firm
ASI Medical, Inc.
Product
SEAC Advanced Dental System, Model A591561 The SEAC Advanced Dental System is a self-contained dental delivery system with a built-in air compressor,vacuum system and water supply. It is designed operate dental handpiece attachments and suction instruments. The SEAC Advanced Dental System can also be upgraded to include a Micadent air abrasion system mounted inside the unit with the handpiece mounted in a holder along side the other handpiece connections. The Micadent is used to perform intra oral air abrasion procedures including etching and preparing surfaces for composite restoration. The SEAC Advanced Dental System is substantially similar in design and functions as the SEAC Mobile Dental System manufactured by ASI Medical, Inc. The Micadent is manufactured by Medidents and will be supplied to ASIas a completed product. The Micadent is substantially the same as it is provided and is attached to an accessory air line inside the SEAC unit for operation. It is intended for use to provide general dental restorative care and hygiene procedures in both traditional dental office settings and/or mobile applications by or under the supervision of a licensed dentist or a hygienist if permitted by applicable law.
Status
Terminated
Initiated
2013-11-05
Posted
2013-12-19
Terminated
2014-08-11
Reason
ASI Medical is initiating a field correction for the SEAC Advanced Dental System because the miniature plastic air tank can potentially rupture.
Root cause
Device Design
Action
ASI Medical sent an Field Correction Action and Safety Notice letter dated November 5, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customers. REQUIRED FIELD CORRECTION Our field correction will consist of sending you a warning label to be applied to the back service access panel of the cart and an ASME safety pop off valve that will open and exhaust if pressure should reach over 100 psi. This safety pop off valve is designed be easily installed in line between the air compressor and air tank in your system. We are requesting your cooperation in making this field correction.
Quantity
337
Distribution
Worldwide Distribution - USA (nationwide) including the countries of Austria, Aruba, Saudi Arabia, Canada, Germany, Spain and West Indies.
Lot, serial or model codes
Code No: 2010, 2020,2120 and 2220 Serial No: 590341- 592066
510(k) numbers
K974487
PMA numbers
not on file
Product code
EIA
Firm FEI
1000160293
Firm city
Centennial
Firm state
CO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register