FDA Medical Device Recalls (openFDA)
Beckman Coulter Inc: COULTER LH 750 Hematology Analyzer.
- Recall number
- Z-0536-05
- Recalling firm
- Beckman Coulter Inc
- Product
- COULTER LH 750 Hematology Analyzer.
- Status
- Terminated
- Initiated
- 2004-12-16
- Posted
- 2005-02-10
- Terminated
- 2012-04-26
- Reason
- A rare risk of sample misidentification exists with Coulter LH 750 Hematology Analyzers. The printout results from the instrument are affected; however, the results stored in the workstation's database are not affected.
- Root cause
- Other
- Action
- Letter 12/20/04 A rare risk of sample misidentification exists with Coulter LH 750 Hematology Analyzers. The printout results from the instrument are affected; however, the results stored in the workstation''s database are not affected. If the LH750 instrument displays a pop-up winder titled Database Error (XXXXXX) and text similar to (Microsoft) (ODBC SQL...) As soon as possible, call Beckman Coulter Customer Service, 800-526-7694) or contact your local rep, to schedule a service call. Do not acknowledge the error. Instead, restart the entire system by completely powering it OFF and then ON. Or, if the LH750 Workstation experiences a change in performance, such as screen display freezes, slow response time, or requires frequent resets to continue processing. As soon as possible, call Beckman Coulter Customer Service, 800-526-7694) or contact your local rep, to schedule a service call. Verify the accuracy of all patient reports for the samples until service confirms the performance of system.
- Quantity
- 1,257
- Distribution
- US and Canada
- Lot, serial or model codes
- Part Number 6605632
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- GKZ
- Firm FEI
- 2050012
- Firm city
- Brea
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28