Skip to the content

FDA Medical Device Recalls (openFDA)

Beckman Coulter Inc: COULTER LH 750 Hematology Analyzer.

Recall number
Z-0536-05
Recalling firm
Beckman Coulter Inc
Product
COULTER LH 750 Hematology Analyzer.
Status
Terminated
Initiated
2004-12-16
Posted
2005-02-10
Terminated
2012-04-26
Reason
A rare risk of sample misidentification exists with Coulter LH 750 Hematology Analyzers. The printout results from the instrument are affected; however, the results stored in the workstation's database are not affected.
Root cause
Other
Action
Letter 12/20/04 A rare risk of sample misidentification exists with Coulter LH 750 Hematology Analyzers. The printout results from the instrument are affected; however, the results stored in the workstation''s database are not affected. If the LH750 instrument displays a pop-up winder titled Database Error (XXXXXX) and text similar to (Microsoft) (ODBC SQL...) As soon as possible, call Beckman Coulter Customer Service, 800-526-7694) or contact your local rep, to schedule a service call. Do not acknowledge the error. Instead, restart the entire system by completely powering it OFF and then ON. Or, if the LH750 Workstation experiences a change in performance, such as screen display freezes, slow response time, or requires frequent resets to continue processing. As soon as possible, call Beckman Coulter Customer Service, 800-526-7694) or contact your local rep, to schedule a service call. Verify the accuracy of all patient reports for the samples until service confirms the performance of system.
Quantity
1,257
Distribution
US and Canada
Lot, serial or model codes
Part Number 6605632
510(k) numbers
not on file
PMA numbers
not on file
Product code
GKZ
Firm FEI
2050012
Firm city
Brea
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register