Skip to the content

FDA Medical Device Recalls (openFDA)

Glycar SA Pty., Ltd.: SJM Pericardial Patch, with EnCap Technology 5x10 cm, Model Number REF C0510; PATCH, PLEDGET AND INTRACARDIAC, PETP, PTFE, POLYPROPYLENE

Recall number
Z-0532-2026
Recalling firm
Glycar SA Pty., Ltd.
Product
SJM Pericardial Patch, with EnCap Technology 5x10 cm, Model Number REF C0510; PATCH, PLEDGET AND INTRACARDIAC, PETP, PTFE, POLYPROPYLENE
Status
Open, Classified
Initiated
2025-10-09
Posted
2025-11-25
Terminated
not on file
Reason
The impacted lot may not meet the required tensile strength specification.
Root cause
Process control
Action
Glycar issued a Stock Withdrawal Notification to its sole distributor on 10/09/2025. The notice explained the issue and requested the following: "Required Actions 1. Immediately identify and segregate all affected stock in your possession. 2. Clearly label the identified stock as "WITHDRAWN PRODUCT" DO NOT DISTRIBUTE OR USE. 3. Suspend any further distribution or transfer of the affected lots. 4. Provide written confirmation of the quantity and storage location of all affected stock within five (5) business days of receipt of this notification. 5. Our Quality team will coordinate collection, return, or replacement of affected product as appropriate." For questions or require further clarification, contact: Malcolm Applewhite, malcolm@glycar.co.za On 10/17/2025, the firm issued an URGENT: MEDICAL DEVICE RECALL notice the medical facility via email. This notice explained the issue, potential risk to health, and requested the following: "Actions to be Taken by the Customer / User: 1. Identify and quarantine any affected products with lot number T2502257 in your possession. 2. Cease use of the affected device immediately. 3. Share this notice with applicable personnel within your institution. 4. Abbott will coordinate and arrange for return and replacement of affected units. 5. Complete and return the attached Acknowledgment Form within 5 business days to confirm receipt of this notice." For questions or assistance, contact: Regulatory Affairs Manager, Glycar SA (Pty) Ltd, Malcolm Applewhite, malcolm@glycar.co.za
Quantity
18 units
Distribution
US
Lot, serial or model codes
UDI/DI 05414734007269, Lot Number T2506245, exp. 2028.05.25
510(k) numbers
K963967
PMA numbers
not on file
Product code
DXZ
Firm FEI
3002807088
Firm city
Irene
Firm state
not on file
Firm country
South Africa

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register