NuVasive Specialized Orthopedics, Inc.: MAGEC 1 System
- Recall number
- Z-0528-2022
- Recalling firm
- NuVasive Specialized Orthopedics, Inc.
- Product
- MAGEC 1 System
- Status
- Open, Classified
- Initiated
- 2019-06-25
- Posted
- not on file
- Terminated
- not on file
- Reason
- Safety Alert to all surgeons informing them of post-implantation fractures of an internal metallic component for device manufactured prior to March 26, 2015.
- Root cause
- Device Design
- Action
- Beginning on 06/25/2019, a Field Safety Notice was communicated to all surgeon informing them that post-implantation fracture of an internal metallic component (i.e., locking pin) has been observed in early versions of the MAGEC System Rods that were manufactured prior to March 26, 2015. Fracture of the locking pin may affect the ability of the device to lengthen and may be associated with Titanium wear debris generation and localized tissue discoloration. Recommended User Actions: - NuVasive, Inc. voluntarily issues this Safety Alert to reinforce the Instructions for Use (IFU) of the MAGEC System to clinical users - The IFU should be consulted on an ongoing basis throughout patient treatment with the MAGEC System. - Users should follow the appropriate postoperative procedure to assess the MAGEC System by X-ray imaging whenever the device is adjusted or at a minimum of once every six months. Typically, the device can be adequately visualized, and a fractured locking pin can be detected, on standard anteroposterior X-ray imaging. Figure 1 shows representative X-ray images for comparison of a device with an intact locking pin (A) to a device manufactured prior to March 26, 2015 with a fractured locking pin (B). A fractured locking pin can typically be identified by examining the noted areas to detect the separation of the internal components and/or the extension of an internal component beyond the device housing. - If a fractured locking pin is detected, removal of the device may be indicated. The decision to remove the device should be made by the physician in consultation with the patient and/or family. - If removal is deemed appropriate by the physician, the device and all associated accessories should be removed and the explanted device should be returned to NuVasive. - Clinical users should review the IFU and be aware of the warnings and precautions on an ongoing basis throughout patient treatment with the device. - Patients and/or families should be reminde
- Quantity
- 3,056 devices
- Distribution
- Worldwide
- Lot, serial or model codes
- All MAGEC Systems (Serial/Lot Numbers) manufactured prior to March 26, 2015.
- 510(k) numbers
- ["K140178"]
- PMA numbers
- not on file
- Product code
- PGN
- Firm FEI
- 3006179046
- Firm city
- Aliso Viejo
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28