CME America, LLC: CME America T-Syringe Pump - T34L PCA - Product Usage: infusion of medications or fluids requiring continuous or intermittent delivery at precisely-controlled infusion rates through clinically acceptable routes of administration, including intravenous, subcutaneous, percutaneous, intra-arterial, epidural, in close proximity to nerves, and into an intra-operative site (soft tissue/body cavity/surgical wound site).

Recall number
Z-0527-2021
Recalling firm
CME America, LLC
Product
CME America T-Syringe Pump - T34L PCA - Product Usage: infusion of medications or fluids requiring continuous or intermittent delivery at precisely-controlled infusion rates through clinically acceptable routes of administration, including intravenous, subcutaneous, percutaneous, intra-arterial, epidural, in close proximity to nerves, and into an intra-operative site (soft tissue/body cavity/surgical wound site).
Status
Terminated
Initiated
2020-07-30
Posted
not on file
Terminated
2022-03-08
Reason
Multiple reasons: 1) Due to wear and tear of pump motor block mechanism - may cause under infusion situations without the pump alarm being generated. 2) Due to cracks, pump hardware may be damaged by fluid used during the cleaning process.
Root cause
Other
Action
On July 30, 2020, the firm sent an "URGENT MEDICAL DEVICE RECALL CORRECTION" Notification Letter via FedEx to customers informing them of two issues: Issue 1: The motor block mechanism may be affected over time by normal wear and tear which may cause the device to stop without alarms being generated. Issue 2: Due to routine cleaning and disinfection, there is a potential for fluid to ingress into the pump and over time, could result in damage to the pump components leading to inoperability of the device. For this issue the device will stop and will alarm. The Recalling Firm is instructing customers to: For Issue 1: Inspect the lead screw to detect the presence of white plastic debris. Inspection must be performed prior to each use of the affected pumps. If signs of wear and tear are observed, discontinue use of the pump and contact the Recalling Firm for product removal. For Issue 2: Follow the Manufacturer Recommended Cleaning instructions in Appendix A and the Technical Service Manual instructions for preventive maintenance provided in Appendix B of the attached Customer Letter to limit the cumulative effects of fluid ingress over time. Immediate Actions to Take: 1) Please refer to the attached customer communication regarding this recall action. 2) As the Recalling Firm plans to notify all affected customers directly, customers are asked to identify customers within their distribution network that purchased the affected product, as defined in the recall notification. Email an excel file listing of all customers to BDRC12@bd.com within 72 hours of receipt of this letter so that the Recalling Firm may initiate customer notification. 3) Complete the attached Distributor Response Form and return to the Recalling Firm's contact noted on the form whether or not the customers have any of the impacted material so that the Recalling Firm may acknowledge the receipt of this notification. 4) Contact the Recalling Firm to arrange for the return of all in sto
Quantity
Total pumps = 4,182 (all models)
Distribution
Worldwide distribution - U.S. Nationwide distribution including in the states of AZ, AK, CA, CO, CT, FL, ID, IL, KS, LA, MA, MI, MO, MT, NC, NE, NV, Oh, OK, OR, PA, TN, TX, WA, and WI. The country of Canada.
Lot, serial or model codes
Model T34L PCA - Part Number(s) (UDI Codes) 100-106PSL (UDI Code: B101100106PSL0) Serial Number: 26486 26485 26484 S93716  100-106PSLM (retired prior to UDI)
510(k) numbers
["K080954"]
PMA numbers
not on file
Product code
FRN
Firm FEI
3006967710
Firm city
Golden
Firm state
CO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register